Barber D C, Brown B H
Sheffield University and Health Authority, Department of Medical Physics and Clinical Engineering, Royal Hallamshire Hospital, UK.
Clin Phys Physiol Meas. 1988;9 Suppl A:101-4. doi: 10.1088/0143-0815/9/4a/017.
Reconstruction of electrical impedance images using the filtered back projection method of Barber and Brown makes several important assumptions about the object being imaged. These are principally that the object has a circular boundary, is two-dimensional and of impedance close to uniform, and has electrodes equally spaced on its boundary. In practice few of these assumptions are met, yet the method appears to give sensible and useful images. This paper looks at errors of reconstruction produced by non-ideal placement of the electrodes and shows that the reconstruction method is insensitive to such placement errors.