Department of Medical Physics, University of Wisconsin-Madison, 1111 Highland Avenue, Madison, WI, 53705, USA.
Department of Radiology, University of Wisconsin-Madison, 600 Highland Avenue, Madison, WI, 53792, USA.
Med Phys. 2020 Sep;47(9):4356-4362. doi: 10.1002/mp.14297. Epub 2020 Jun 23.
A major technical obstacle to bringing x-ray multicontrast (i.e., attenuation, phase, and dark-field) imaging methodology to clinical use is the prolonged data acquisition time caused by the phase stepping procedure. The purpose of this work was to introduce a fast acquisition with seamless stage translation (FASST) technique to a prototype multicontrast breast imaging system for reduced image acquisition time that is clinically acceptable.
The prototype system was constructed based on a Hologic full-field digital mammography + digital breast tomosynthesis combination system. During each FASST acquisition process, a motorized stage holding a diffraction grating travels continuously with a constant velocity, and a train of 15 short x-ray pulses (35 ms each) was delivered by using the Zero-Degree Tomo mode of the Hologic system. Standard phase retrieval was applied to the 15 subimages without spatial interpolation to avoid spatial resolution loss. The method was evaluated using a physical phantom, a bovine udder specimen, and a freshly resected mastectomy specimen. The FASST technique was experimentally compared with single-shot acquisition methods and the standard phase stepping method.
The image acquisition time of the proposed method is 3.7 s. In comparison, conventional phase stepping took 105 s using the same prototype imaging system. The mean glandular dose of both methods was matched at 1.3 mGy. No artifacts or spatial resolution loss was observed in images produced by FASST. In contrast, the single-shot methods led to spatial resolution loss and residual moiré artifacts.
The FASST technique reduces the data acquisition time of the prototype multicontrast x-ray breast imaging system to 3.7 s, such that it is comparable to a clinical digital breast tomosynthesis exam.
将 X 射线多对比(即衰减、相位和暗场)成像方法应用于临床的主要技术障碍是相位步进过程导致的数据采集时间延长。本研究的目的是为原型多对比乳腺成像系统引入快速采集与无缝台架平移(FASST)技术,以减少图像采集时间,使其达到临床可接受的程度。
原型系统基于 Hologic 全数字乳腺摄影+数字乳腺断层合成组合系统构建。在每次 FASST 采集过程中,一个装有衍射光栅的电动台架以恒定速度连续移动,系统的零角度断层(Zero-Degree Tomo)模式下会连续发出 15 个短 X 射线脉冲(每个 35ms)。标准的相位恢复应用于 15 个子图像,无需空间插值以避免空间分辨率损失。该方法使用物理体模、牛乳房标本和新鲜切除的乳房切除术标本进行了评估。FASST 技术与单次采集方法和标准相位步进方法进行了实验比较。
所提出方法的图像采集时间为 3.7s。相比之下,使用相同的原型成像系统,传统的相位步进需要 105s。两种方法的平均腺体剂量匹配在 1.3mGy。FASST 产生的图像中没有观察到伪影或空间分辨率损失。相比之下,单次采集方法导致空间分辨率损失和残余莫尔伪影。
FASST 技术将原型多对比 X 射线乳腺成像系统的数据采集时间缩短至 3.7s,使其与临床数字乳腺断层合成检查相当。