Aurich Konstanze, Fregin Bob, Palankar Raghavendra, Wesche Jan, Hartwich Oliver, Biedenweg Doreen, Nguyen Thi-Huong, Greinacher Andreas, Otto Oliver
Transfusionsmedizin, Universitätsmedizin Greifswald, 17475 Greifswald, Germany.
Lab Chip. 2020 Jun 30;20(13):2306-2316. doi: 10.1039/d0lc00258e.
Without cellular blood products such as platelet concentrates (PC), red blood cell concentrates (RCC), and hematopoietic stem cells (HPSC) modern treatments in medicine would not be possible. An unresolved challenge is the assessment of their quality with minimal cell manipulation. Minor changes in production, storage conditions, or blood bag composition may impact cell function, which can have important consequences on product integrity. This is especially relevant for personalized medicine, such as autologous T-cell therapy. Today a robust methodology that globally determines cell status directly before transfusion or transplantation is lacking. We demonstrate that measuring viscoelastic characteristics of peripheral blood cells using real-time deformability cytometry (RT-DC) provides comprehensive information on product quality, which is not accessible using conventional quality control tests. In addition, RT-DC requires few cells, a minimal sample volume and has a rapid turnaround time. We compared RT-DC to standard in vitro quality assays assessing: i) PC after storage at 4 °C and room temperature; ii) magnetic nanoparticle labeled platelets; iii) RCC stored in blood bags with different plasticizers; iv) RCC after gamma irradiation; and v) HPSC after cryopreservation with 5% or 10% dimethyl sulfoxide, respectively. Additionally, we evaluated the engraftment time of patients' platelets and leukocytes after transplantation of HPSC products. Our results demonstrate that label-free mechano-phenotyping can be used as a potential biomarker for quality assessment of cell-based pharmaceutical products.
没有诸如血小板浓缩物(PC)、红细胞浓缩物(RCC)和造血干细胞(HPSC)等细胞血液制品,现代医学治疗将无法实现。一个尚未解决的挑战是在尽量减少细胞操作的情况下评估它们的质量。生产、储存条件或血袋成分的微小变化可能会影响细胞功能,这可能对产品完整性产生重要影响。这对于个性化医疗,如自体T细胞疗法尤其重要。如今,缺乏一种在输血或移植前直接全面确定细胞状态的可靠方法。我们证明,使用实时变形性细胞术(RT-DC)测量外周血细胞的粘弹性特征可提供有关产品质量的全面信息,而这是传统质量控制测试无法获得的。此外,RT-DC所需细胞数量少、样本体积小且周转时间快。我们将RT-DC与标准体外质量检测方法进行了比较,检测对象包括:i)在4°C和室温下储存后的PC;ii)磁性纳米颗粒标记的血小板;iii)储存在含有不同增塑剂血袋中的RCC;iv)γ射线照射后的RCC;以及v)分别用5%或10%二甲基亚砜冷冻保存后的HPSC。此外,我们评估了HPSC产品移植后患者血小板和白细胞的植入时间。我们的结果表明,无标记机械表型分析可作为基于细胞的药品质量评估的潜在生物标志物。