Department of Periodontology and Oral Surgery, University Hospital of Liège, Liège, Belgium.
Department of Periodontology, Faculty of Dental Medicine, Monastir, Tunisia.
Clin Oral Implants Res. 2020 Sep;31(9):814-824. doi: 10.1111/clr.13624. Epub 2020 Jun 29.
Implants with a triangular neck were recently introduced to limit peri-implant bone loss. The primary objective of this randomized controlled trial was to compare peri-implant bone changes of circular versus triangular cross-section neck implants 1 year after loading. The secondary objectives were to assess buccal hard tissue thickness changes, Pink Esthetic Score (PES), and patient satisfaction.
Thirty four patients requiring replacement of the single, intercalated missing tooth of healed site for at least 4 months in the posterior maxilla were randomized into 2 groups according to the type of implant. Immediately after surgery and 1 year after final restoration, a cone beam CT (CBCT) was performed to assess proximal bone remodeling and buccal bone thickness. Peri-implant soft tissue health, PES, and patient-reported outcome measures (PROMs) were recorded.
No implant loss occurred within the follow-up period. The mean ± SD peri-implant proximal bone loss 1 year after loading was 0.22 ± 0.30 mm for triangular and 0.42 ± 0.67 mm for circular implants necks (p = .25). Peri-implant bone loss exceeding 2 mm was observed in a single implant in the circular neck group. Buccal bone thickness remained stable and did not differ different between the 2 groups. The peri-implant soft tissue health, PES, and patient satisfaction were also comparable.
Within the limitations of the present study, patient clinical and radiographic outcomes did not differ between triangular and circular cross-section neck implants in the posterior maxilla.
最近引入了具有三角形颈部的种植体,以限制种植体周围骨的丧失。本随机对照试验的主要目的是比较加载后 1 年圆形和三角形颈部种植体的种植体周围骨变化。次要目的是评估颊侧硬组织厚度变化、Pink 美学评分(PES)和患者满意度。
34 名患者需要在前磨牙区上颌后部更换单个、间隔缺失的牙齿,这些牙齿在愈合部位至少缺失 4 个月。根据种植体类型将患者随机分为两组。在手术后立即和最终修复后 1 年,进行锥形束 CT(CBCT)检查,以评估近侧骨重塑和颊侧骨厚度。记录种植体周围软组织健康、PES 和患者报告的结果测量(PROMs)。
在随访期间,没有发生种植体丢失。加载后 1 年近侧种植体骨丢失的平均值 ± 标准偏差为三角形颈种植体为 0.22 ± 0.30mm,圆形颈种植体为 0.42 ± 0.67mm(p=0.25)。在圆形颈种植体组中,单个种植体出现超过 2mm 的种植体周围骨丢失。颊侧骨厚度保持稳定,两组之间无差异。种植体周围软组织健康、PES 和患者满意度也相似。
在本研究的限制范围内,后磨牙区三角形和圆形颈部种植体的患者临床和影像学结果无差异。