Analysis and Testing of a Suitable Compatible Electrode's Material for Continuous Measurement of Glucose Concentration.

机构信息

Department of Cybernetic and Biomedical Engineering, VŠB-Technical University of Ostrava, 17, listopadu 2172/15, 708 00 Ostrava-Poruba, Czech Republic.

Malaysia-Japan International Institute of Technology, Universiti Teknologi Malaysia Kuala Lumpur, Jalan Sultan Yahya Petra, Kuala Lumpur 54100, Malaysia.

出版信息

Sensors (Basel). 2020 Jun 30;20(13):3666. doi: 10.3390/s20133666.

Abstract

The subject of the submitted work is the proposal of electrodes for the continual measurement of the glucose concentration for the purpose of specifying further hemodynamic parameters. The proposal includes the design of the electronic measuring system, the construction of the electrodes themselves and the functionality of the entire system, verified experimentally using various electrode materials. The proposed circuit works on the basis of micro-ammeter measuring the size of the flowing electric current and the electrochemical measurement method is used for specifying the glucose concentration. The electrode system is comprised of two electrodes embedded in a silicon tube. The solution consists of the measurement with three types of materials, which are verified by using three solutions with a precisely given concentration of glucose in the form of a mixed solution and enzyme glucose oxidase. For the testing of the proposed circuit and the selection of a suitable material, the testing did not take place on measurements in whole blood. For the construction of the electrodes, the three most frequently used materials for the construction of electrodes used in clinical practice for sensing biopotentials, specifically the materials Ag/AgCl, Cu and Au, were used. The performed experiments showed that the material Ag/AgCl, which had the greatest sensitivity for the measurement even without the enzyme, was the most suitable material for the electrode. This conclusion is supported by the performed statistical analysis. On the basis of the testing, we can come to the conclusion that even if the Ag/AgCl electrode appears to be the most suitable, showing high stability, gold-plated electrodes showed stability throughout the measurement similarly to Ag/AgCl electrodes, but did not achieve the same qualities in sensitivity and readability of the measured results.

摘要

提交工作的主题是提出用于连续测量葡萄糖浓度以进一步指定血液动力学参数的电极。该提案包括电子测量系统的设计、电极本身的结构和整个系统的功能,使用各种电极材料进行了实验验证。所提出的电路基于微安培计工作,该安培计测量流动电流的大小,并且使用电化学测量方法来指定葡萄糖浓度。电极系统由嵌入在硅管中的两个电极组成。该解决方案包括使用三种材料进行测量,通过使用三种混合溶液和酶葡萄糖氧化酶的精确给定浓度的葡萄糖溶液来验证这三种材料。为了测试所提出的电路和选择合适的材料,测试未在全血测量中进行。为了构建电极,使用了在临床实践中用于感测生物电势的电极构造中最常用的三种材料,即 Ag/AgCl、Cu 和 Au。所进行的实验表明,即使没有酶,Ag/AgCl 材料对测量具有最大的灵敏度,是电极的最适合材料。该结论得到了进行的统计分析的支持。基于测试,我们可以得出结论,即使 Ag/AgCl 电极似乎是最合适的,具有高稳定性,但镀金电极在整个测量过程中表现出与 Ag/AgCl 电极相似的稳定性,但在灵敏度和测量结果的可读性方面并未达到相同的质量。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/0ee3/7374362/34e2ca5e6106/sensors-20-03666-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验