Abstract

Automated data processing and quality control of radioimmunoassays offer not only increased speed but also a more thorough and statistically rigorous analysis of results. An external quality assessment scheme for serum thyroxine, triiodothyronine and thyroid stimulating hormone (TSH) assays was performed in five nuclear medicine laboratories in Korea to compare with the assay performances of the World Health Organization Radioimmunoassay Program. The required radioimmunoassay kits were supplied through the International Atomic Energy Agency (IAEA). We have determined the weighted root mean squared error, and variance ratio as the indices of standard curve and also the average batch coefficient of variation (ABCV) as the parameters of response error relationship curve and precision profile. There was a good fit for the triiodothyronine assay, but 3 of 5 laboratories showed possible bad fit in the T4 and TSH assay systems. The ABCV was less than 5 percent for the T3 and T4 assay system, however for the TSH system, only 1 laboratory showed the ABCV value of less than 5 percent. We have also calculated the within batch variation (drift) and between laboratory variations.

摘要

放射免疫分析的自动化数据处理和质量控制不仅提高了速度,还能对结果进行更全面、统计上更严格的分析。韩国的五个核医学实验室开展了一项针对血清甲状腺素、三碘甲状腺原氨酸和促甲状腺激素(TSH)检测的外部质量评估计划,以与世界卫生组织放射免疫分析计划的检测性能进行比较。所需的放射免疫分析试剂盒由国际原子能机构(IAEA)提供。我们确定了加权均方根误差和方差比作为标准曲线的指标,并确定了平均批内变异系数(ABCV)作为反应误差关系曲线和精密度曲线的参数。三碘甲状腺原氨酸检测拟合良好,但在5个实验室中,有3个实验室的T4和TSH检测系统可能拟合不佳。T3和T4检测系统的ABCV小于5%,然而对于TSH系统,只有1个实验室的ABCV值小于5%。我们还计算了批内变异(漂移)和实验室间变异。

相似文献

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验