Suppr超能文献

生物流体中总二氧化碳和氯化物的同步电化学测定(作者译)

[Simultaneous electrochemical determinations of total co2 and chlorides in biological fluids (author's transl)].

作者信息

Bousquet B, Garnier J P, Dreux C

出版信息

Pathol Biol (Paris). 1977 May;25(5):353-8.

PMID:327416
Abstract

We thought it was interesting to study the results provided by an instrument intended for chlorides and total CO2 plasmatic determinations using specific electrodes. The main qualities observed are the following: -Chloride linearity 10 up to 300 mmol/I. -Total CO2 linearity 5 up to 50 mmol/I. The within run precision for chloride is 0,74% CV and for total CO2 is 1,56% CV. The between day precision is excellent: CI--0,95% CV. Total CO2 1,54% CV. --No contamination at all. --Throughput: 55 samples per hour. --Good correlation with the routine methods used in our laboratory CI--r = 0,90. Total CO2 r = 0,98. --No interferences at all by compounds such as hemoglobulin, bilirubin. --Possible measurement with urine, CSF, gastric fluid, ascites and pleural punctures and sweat. This instrument seems well adapted for pediatric, resuscitation and "stat" setting in a clinical laboratory.

摘要

我们认为研究一种使用特定电极进行血浆氯化物和总二氧化碳测定的仪器所提供的结果很有意思。观察到的主要特性如下:

  • 氯化物线性范围为10至300 mmol/L。

  • 总二氧化碳线性范围为5至50 mmol/L。

  • 氯化物批内精密度为0.74% CV,总二氧化碳批内精密度为1.56% CV。

  • 日间精密度极佳:氯化物为0.95% CV,总二氧化碳为1.54% CV。

  • 完全无污染。

  • 通量:每小时55个样本。

  • 与我们实验室使用的常规方法相关性良好:氯化物r = 0.90,总二氧化碳r = 0.98。

  • 不受血红蛋白、胆红素等化合物的干扰。

  • 可用于尿液、脑脊液、胃液、腹水、胸腔穿刺液和汗液的检测。该仪器似乎非常适合临床实验室的儿科、复苏和“急诊”检测。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验