Abstract

PURPOSE

Correct commissioning of treatment planning systems (TPSs) is important for reducing treatment failure events. There is currently no comprehensive and robust methodology available for TPS commissioning in modern brachytherapy. This review aimed to develop a comprehensive template for commissioning modern 3D-image-based brachytherapy TPSs for high dose rate (HDR) gynaecological applications.

METHODS

The literature relevant to TPS commissioning, including both external beam radiation therapy (EBRT) and brachytherapy, as well as guidelines by the International Atomic Energy Agency (IAEA), the American Association of Physicists in Medicine (AAPM), and the European Society for Radiotherapy and Oncology (ESTRO) were searched, studied and appraised. The applied relevant EBRT TPS commissioning tests were applied to brachytherapy. The developed template aimed to cover all dosimetric and non-dosimetric issues.

RESULTS

The essential commissioning items could be categorized into six parts: geometry, dose calculation, plan evaluation tools, plan optimization, TPS output, and end-to-end verification. The final template consists of 43 items. This paper presents the purpose and role of each test, as well as tolerance limits, to facilitate the use of the template.

CONCLUSION

The information and recommendations available in a collection of publications over many years have been reviewed in order to develop a comprehensive template for commissioning complex modern 3D-image-based brachytherapy TPSs for HDR gynaecological applications. The up-to-date and concise information contained in the template can aid brachytherapy physicists during TPS commissioning as well as devising a regular quality assurance program and allocation of time and resources.

摘要

目的

正确调试治疗计划系统(TPS)对于减少治疗失败事件非常重要。目前,现代近距离放射治疗中尚无全面、稳健的 TPS 调试方法。本研究旨在为现代基于 3D 图像的 HDR 妇科近距离放射治疗 TPS 开发一个全面的调试模板。

方法

搜索、研究和评估了与 TPS 调试相关的文献,包括外束放射治疗(EBRT)和近距离放射治疗,以及国际原子能机构(IAEA)、美国医学物理学家协会(AAPM)和欧洲放射治疗和肿瘤学学会(ESTRO)的指南。将应用的相关 EBRT TPS 调试测试应用于近距离放射治疗。开发的模板旨在涵盖所有剂量学和非剂量学问题。

结果

基本调试项目可分为六个部分:几何形状、剂量计算、计划评估工具、计划优化、TPS 输出和端到端验证。最终模板包括 43 个项目。本文介绍了每个测试的目的和作用,以及容限限制,以方便模板的使用。

结论

本研究综合了多年来发表的多篇文献中的信息和建议,为 HDR 妇科应用的复杂现代基于 3D 图像的近距离放射治疗 TPS 调试开发了一个全面的模板。模板中包含的最新、简洁的信息可以帮助近距离放射治疗物理学家在 TPS 调试以及制定常规质量保证计划和分配时间和资源时提供帮助。