J Med Syst. 2020 Aug 18;44(10):174. doi: 10.1007/s10916-020-01642-2.
The aim of this study is to present the SurgeForce system, a tissue handling training device for analysis of dynamic force applied to the tissue and objective assessment of basic surgical skills during the suture process. The SurgeForce system consists of a mechanical base formed by two platforms joint with three stainless steel springs and a three axial digital accelerometer attached to the upper platform, which detects the dynamic force caused by a surgeon when performing a suture task over a synthetic tissue pad. Accelerometer data is sent to a control unit where preprocessing to transform the raw data into a force signal is done, and then, the force signal is sent to a computer application, which register the force exerted over the synthetic tissue pad. For validation, 17 participants (6 surgeons and 11 medical students) performed three simple interrupted sutures with knot tying using the SurgeForce system. Ten force-based metrics were proposed to evaluate their performance during the suturing task. Results of the validation showed statistical differences in 8 of 10 force-based parameters for assessment of basic surgical skills during the suture task. The SurgeForce system demonstrated its capacity to differentiate force-based performance of surgeons and medical students. The SurgeForce system has been successfully validated. This system was able to distinguish force performance between experts and novices, showing its potential to distinguish surgeons with basic suture skills from those who are not yet prepared.
本研究旨在介绍 SurgeForce 系统,这是一种组织处理训练设备,用于分析缝合过程中施加于组织的动态力,并对基本手术技能进行客观评估。SurgeForce 系统由两个平台组成,通过三个不锈钢弹簧连接,一个三轴数字加速度计连接在上平台上,用于检测外科医生在合成组织垫上进行缝合任务时产生的动态力。加速度计数据被发送到控制单元,在那里对原始数据进行预处理,将其转换为力信号,然后将力信号发送到计算机应用程序,该程序记录在合成组织垫上施加的力。为了验证,17 名参与者(6 名外科医生和 11 名医学生)使用 SurgeForce 系统进行了三次带有打结的简单间断缝合。提出了十个基于力的指标来评估他们在缝合任务中的表现。验证结果显示,在评估缝合任务中基本手术技能的 10 个基于力的参数中,有 8 个存在统计学差异。SurgeForce 系统证明了其区分外科医生和医学生基于力的表现的能力。SurgeForce 系统已成功验证。该系统能够区分专家和新手之间的力性能,表明其有潜力区分具有基本缝合技能的外科医生和尚未准备好的外科医生。