后房型有晶状体眼人工晶状体植入术后房角开放参数对角膜内皮细胞密度和眼压的影响

Effect of Angle Opening Parameters on Corneal Endothelial Cell Density and Intraocular Pressure after Posterior Chamber Phakic Intraocular Lens Implantation.

作者信息

Kamiya Kazutaka, Ando Wakako, Tsujisawa Tatsuhiko, Takahashi Masahide, Shoji Nobuyuki

机构信息

Visual Physiology, School of Allied Health Sciences, Kitasato University, Kanagawa 252-0373, Japan.

Department of Ophthalmology, School of Medicine, Kitasato University, Kanagawa 252-0374, Japan.

出版信息

J Clin Med. 2020 Aug 21;9(9):2704. doi: 10.3390/jcm9092704.

Abstract

This study aimed to investigate the time courses of angle opening parameters and the relationships of these with the corneal endothelial cell density (ECD) and the intraocular pressure (IOP) after posterior chamber phakic intraocular lens (Visian ICL, STAAR Surgical) implantation. We evaluated 116 eyes of 59 consecutive patients (mean age ± standard deviation, 34.0 ± 8.8 years) who underwent V5 implantable collamer lens (ICL) implantation. Preoperatively and 1 day, 1 week, and 1, 3, and 18 months postoperatively, we quantitatively measured the angle opening distance at 500 µm (AOD500), the trabecular-iris space area (TISA500), and the trabecular iris angle (TIA500), using anterior segment optical coherence tomography (CASIA 2, Tomey), and assessed the relationships of these measurements with ECD and IOP in ICL-implanted eyes. All angle parameters (AOD500, TISA500, and TIA500) significantly decreased 1 day postoperatively but remained stable thereafter. At 18 months postoperatively, we found no significant correlations of the angle parameters with ECD (Pearson correlation coefficient = -0.108, = 0.249 for AOD500; = -0.162, = 0.083 for TISA500; = -0.022, = 0.815 for TIA500) or between the angle parameters and IOP ( = -0.106, = 0.256 for AOD500; = -0.021, = 0.826 for TISA500; = -0.018, = 0.850 for TIA500). The angle opening metrics significantly decreased immediately after ICL implantation but remained stable thereafter. Narrowing of the angle did not significantly affect ECD or IOP in ICL-implanted eyes during the 18-month postoperative period.

摘要

本研究旨在探讨后房型有晶状体眼人工晶状体(Visian ICL,STAAR Surgical公司)植入术后房角开放参数的时间进程,以及这些参数与角膜内皮细胞密度(ECD)和眼压(IOP)之间的关系。我们评估了59例连续患者的116只眼(平均年龄±标准差,34.0±8.8岁),这些患者均接受了V5可植入式胶原晶状体(ICL)植入术。在术前以及术后1天、1周、1个月、3个月和18个月,我们使用眼前节光学相干断层扫描(CASIA 2,Tomey公司)定量测量了500 µm处的房角开放距离(AOD500)、小梁-虹膜空间面积(TISA500)和小梁虹膜角(TIA500),并评估了这些测量值与ICL植入眼的ECD和IOP之间的关系。所有房角参数(AOD500、TISA500和TIA500)在术后1天均显著降低,但此后保持稳定。术后18个月,我们发现房角参数与ECD之间无显著相关性(AOD500的Pearson相关系数 = -0.108,P = 0.249;TISA500的Pearson相关系数 = -0.162,P = 0.083;TIA500的Pearson相关系数 = -0.022,P = 0.815),房角参数与IOP之间也无显著相关性(AOD500的Pearson相关系数 = -0.106,P = 0.256;TISA500的Pearson相关系数 = -0.021,P = 0.826;TIA500的Pearson相关系数 = -0.018,P = 0.850)。ICL植入后房角开放指标立即显著降低,但此后保持稳定。在术后18个月期间,房角变窄对ICL植入眼的ECD或IOP无显著影响。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/d1cc/7565728/a9e8aa946fdc/jcm-09-02704-g001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验