A Novel Automated Visual Acuity Test Using a Portable Head-mounted Display.

机构信息

Yong Loo Lin School of Medicine, National University of Singapore, Singapore, Singapore.

Department of Ophthalmology, National University Hospital, Singapore, Singapore.

出版信息

Optom Vis Sci. 2020 Aug;97(8):591-597. doi: 10.1097/OPX.0000000000001551.

Abstract

SIGNIFICANCE

We developed a head-mounted display (HMD) as an automated way of testing visual acuity (VA) to increase workplace efficiency. This study raises its potential utility and advantages, analyzes reasons for its current limitations, and discusses areas of improvement in the development of this device.

PURPOSE

Manual VA testing is important but labor-intensive in ophthalmology and optometry clinics. The purpose of this exploratory study is to assess the performance and identify potential limitations of an automated HMD for VA testing.

METHODS

Sixty patients from National University Hospital, Singapore, were enrolled in a prospective observational study. The HMD was constructed based on the Snellen chart, with single optotypes displayed at a time. Each subject underwent VA testing of both eyes with the manual Snellen chart tested at 6 m from the subject and the HMD.

RESULTS

Fifty-three subjects were included in the final analysis, with an incompletion rate of 11.7% (n = 7). The mean difference in estimated acuity between the HMD and Snellen chart was 0.05 logMAR. However, 95% limits of agreement were large at ±0.33 logMAR. The HMD overestimated vision in patients with poorer visual acuities. In detecting VA worse than 0.30 logMAR (6/12), sensitivity was 63.6% (95% confidence interval, 0.31 to 0.89%), and specificity was 81.0% (95% confidence interval, 0.66 to 0.91%). No significant correlation existed between mean difference and age (r = -0.15, P = .27) or education level (r = 0.04, P = .76).

CONCLUSIONS

Advantages of our novel HMD technology include its fully automated nature and its portability. However, the device in its current form is not ready for widespread clinical use primarily because of its low accuracy, which is limited by both technical and user factors. Future studies are needed to improve its accuracy and completion rate and to evaluate for test-retest reliability in a larger population.

摘要

意义

我们开发了一种头戴式显示器 (HMD),作为一种自动测试视力 (VA) 的方法,以提高工作场所的效率。本研究提出了其潜在的效用和优势,分析了其目前限制的原因,并讨论了该设备开发中的改进领域。

目的

手动 VA 测试在眼科和验光配镜诊所中很重要,但很耗费劳力。本探索性研究的目的是评估一种自动 HMD 进行 VA 测试的性能并确定其潜在限制。

方法

新加坡国立大学医院的 60 名患者参与了这项前瞻性观察研究。HMD 是基于 Snellen 图表构建的,一次显示单个视标。每位受试者均接受双眼 VA 测试,使用手动 Snellen 图表在距离受试者 6 米处进行测试,以及 HMD 进行测试。

结果

最终分析纳入了 53 名受试者,完成率为 11.7%(n = 7)。HMD 和 Snellen 图表估计视力之间的平均差异为 0.05 logMAR。然而,95%的一致性界限为±0.33 logMAR。HMD 高估了视力较差的患者的视力。在检测视力低于 0.30 logMAR(6/12)时,灵敏度为 63.6%(95%置信区间,0.31 至 0.89%),特异性为 81.0%(95%置信区间,0.66 至 0.91%)。平均差异与年龄(r = -0.15,P =.27)或教育水平(r = 0.04,P =.76)之间无显著相关性。

结论

我们新型 HMD 技术的优势包括其全自动性质及其便携性。然而,该设备目前的形式还不能广泛用于临床,主要是因为其准确性较低,这受到技术和用户因素的限制。需要进一步的研究来提高其准确性和完成率,并在更大的人群中评估其测试-再测试的可靠性。