采用骨诱导钻孔方案在不同密度的聚亚安酯块中增加种植体初期稳定性。

Implant primary stability with an osteocondensation drilling protocol in different density polyurethane blocks.

机构信息

Department of Medical, Oral and Biotechnological Sciences, University "G. D'Annunzio" of Chieti-Pescara, Chieti, Italy.

Department of Interdisciplinary Medicine, University of Bari "Aldo Moro", Bari, Italy.

出版信息

Comput Methods Biomech Biomed Engin. 2021 Jan;24(1):14-20. doi: 10.1080/10255842.2020.1806251. Epub 2020 Aug 25.

Abstract

The implant primary stability is a fundamental condition for avoiding implant micro-motions that might result in fibrous encapsulation; its achievement is facilitated by macro- and micro-geometry of the implant, and by the bone density and architecture at the intended implant site. The aim was to evaluate an osteocondensing drilling protocol for dental implant positioning compared to standard protocol on polyurethane block. A total of 40 implants, 20 for each osteocondensing group (Test) and 20 for standard drilling group (Control), were positioned. Insertion torque (IT), removal torque (RT), and Periotest were measured. A IT, RT, and Periotest significant difference was present in favor of the implant osteocondensing protocol. The results suggested that osteocondensing protocol represents a useful technique for implant placement in poor density bone.

摘要

种植体初期稳定性是避免种植体微动从而导致纤维包裹的基本条件;其实现得益于种植体的宏观和微观几何形状,以及预期种植部位的骨密度和结构。本研究旨在评估骨压缩钻技术与标准技术在聚亚安酯块上定位牙种植体的效果。共定位 40 个种植体,每组 20 个,骨压缩组(试验)和标准钻孔组(对照)。测量了插入扭矩(IT)、去除扭矩(RT)和 Periotest。IT、RT 和 Periotest 有显著差异,有利于种植体骨压缩技术。结果表明,骨压缩技术是在骨密度低的情况下进行种植体放置的一种有效技术。