Electromagnetic interference effect of dental equipment on cardiac implantable electrical devices: A systematic review.

机构信息

Faculty of Stomatology, The First Clinical Medical School, Tongji Medical College, Huazhong University of Science and Technology, Hubei, China.

Department of Cardiology, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.

出版信息

Pacing Clin Electrophysiol. 2020 Dec;43(12):1588-1598. doi: 10.1111/pace.14051. Epub 2020 Sep 12.

Abstract

BACKGROUND

The electromagnet interference (EMI) effect resulting from using dental equipment near cardiovascular implantable electronic devices (CIEDs) is controversial based on in vitro and in vivo studies. We aimed to summarize the available evidence to investigate the safety of using dental equipment on patients with CIEDs.

METHODS

An electronic search was performed in PubMed, Embase, MEDLINE Ovid, and the Cochrane Library for relevant studies published between January 2000 and May 2020. The search strategy centered on terms related to dental devices and CIEDs. Two independent reviewers determined the final inclusion of the studies in the systematic review. The EMI effect was summarized based on different dental instruments detected in in vitro or in vivo studies.

RESULTS

The primary search identified 84 articles, and 18 studies were finally included in this systematic review after exclusions. Most in vitro studies (n = 12) reported background noise or severe EMI affecting CIED function at a close distance from the lead tip or at a high sensitivity setting of CIEDs. In in vivo studies (n = 6), EMI that altered CIED function was not detected at clinical distance and sensitivity settings. The summary, based on electronic apex locators, ultrasonic devices, and electric pulp testers, demonstrated the compatibility of these common dental devices with CIEDs.

CONCLUSIONS

This systematic review indicates that most dental instruments can be used safely in routine dental practice. The EMI effect of dental equipment depends on the exposure distance and lead-related parameters of the CIEDs.

摘要

背景

基于体外和体内研究,使用牙科设备靠近心血管植入式电子设备(CIED)会产生电磁干扰(EMI)效应,这一效应仍存在争议。我们旨在总结现有证据,以调查 CIED 患者使用牙科设备的安全性。

方法

我们在 PubMed、Embase、MEDLINE Ovid 和 Cochrane Library 中进行了电子检索,以查找 2000 年 1 月至 2020 年 5 月期间发表的相关研究。检索策略主要围绕与牙科设备和 CIED 相关的术语。两名独立的综述作者确定了系统评价中最终纳入的研究。根据体外或体内研究中检测到的不同牙科仪器,总结了 EMI 效应。

结果

最初的检索确定了 84 篇文章,经过排除后,最终有 18 项研究纳入了本系统评价。大多数体外研究(n=12)报告称,在距离导联尖端较近或在 CIED 高灵敏度设置下,背景噪声或严重 EMI 会影响 CIED 功能。在体内研究(n=6)中,在临床距离和灵敏度设置下,未检测到改变 CIED 功能的 EMI。基于电子根尖定位仪、超声设备和电活力计的总结表明,这些常见的牙科设备与 CIED 兼容。

结论

本系统评价表明,在常规牙科实践中,大多数牙科仪器都可以安全使用。牙科设备的 EMI 效应取决于 CIED 的暴露距离和导联相关参数。