University College London (UCL), Institute of Ophthalmology, London, UK.
NIHR Moorfields Biomedical Research Centre, London, London, UK.
Transl Vis Sci Technol. 2020 Jul 13;9(8):18. doi: 10.1167/tvst.9.8.18. eCollection 2020 Jul.
To describe, validate, and provide preliminary normative data for an open-source eye-movement perimeter (Eyecatcher).
Visual field testing was performed monocularly in 64 normally sighted young adults, using (i) a Humphrey Field Analyzer (HFA) and (ii) the novel Eyecatcher procedure. Eyecatcher used a remote eye-tracker to position stimuli relative to the current point of fixation, and observers responded by looking towards the stimulus. In both tests, Goldman III stimuli were sampled from a 24-2 grid, and were presented against a 10 cd/m background. Participants completed each test twice to assess test-retest repeatability.
Mean Sensitivity (MS) did not differ between Eyecatcher and the HFA ( = 0.086), and both tests exhibited similar test-retest repeatability (CoR = ±1.86 dB; CoR = ±1.95 dB). Eyecatcher was also able to detect changes in sensitivity across the normal visual field (the "Hill of Vision"), and could differentiate the physiological blind spot from adjacent retinal locations. Mean sensitivities and 95% limits of agreement are described for each pointwise location.
Eyecatcher can use eye movements to assess visual fields in young, normally sighted adults. In such observers, it provides results similar to the current gold standard clinical device (HFA).
Given further development, eye movement perimeters such as Eyecatcher could be particularly useful for individuals unable to perform traditional perimetric assessments, such as young children or stroke patients. Full technical details and information on how to freely acquire the source code are included.
描述、验证和提供开源眼动仪(Eyecatcher)的初步规范数据。
对 64 名视力正常的年轻成年人进行单眼视野测试,使用(i)Humphrey 视野分析仪(HFA)和(ii)新型 Eyecatcher 程序。Eyecatcher 使用远程眼动追踪器将刺激物相对于当前注视点的位置,观察者通过注视刺激物来做出反应。在这两种测试中,Goldman III 刺激物都从 24-2 网格中取样,并在 10 cd/m 的背景下呈现。参与者完成每个测试两次,以评估测试-重测的可重复性。
Eyecatcher 和 HFA 的平均敏感度(MS)没有差异( = 0.086),并且两种测试都表现出相似的测试-重测可重复性(CoR = ±1.86 dB;CoR = ±1.95 dB)。Eyecatcher 还能够检测到正常视野范围内的敏感度变化(“视野山”),并能够将生理盲点与相邻视网膜位置区分开来。描述了每个点位置的平均敏感度和 95%的一致性界限。
Eyecatcher 可以使用眼球运动来评估视力正常的年轻成年人的视野。在这些观察者中,它提供的结果与当前的金标准临床设备(HFA)相似。
田颖