Department of Trauma and Orthopaedics, Basingstoke and North Hampshire Hospital, Hampshire Hospitals NHS Foundation Trust, Basingstoke, United Kingdom.
Department of Trauma and Orthopaedics, Basingstoke and North Hampshire Hospital, Hampshire Hospitals NHS Foundation Trust, Basingstoke, United Kingdom; Department of Trauma and Orthopaedics, Queen Elizabeth Hospital, Gateshead Health NHS Foundation Trust, Gateshead, United Kingdom.
J Arthroplasty. 2021 Feb;36(2):616-622. doi: 10.1016/j.arth.2020.08.013. Epub 2020 Aug 10.
We assessed the survivorship of a proximally hydroxyapatite coated, double tapered, titanium-alloy femoral stem in a single center, at an average follow up of 12.5 years (10.1-15.8). The majority of stems were inserted as part of a Metal on Metal (MoM) Total Hip Replacement (THR).
Data was collected prospectively in a local database. A retrospective review was performed of all patients undergoing a primary THR with the prosthesis between 2003 and 2010. Primary outcome was revision of the stem for any cause. Analysis was also performed for stem revision for aseptic loosening, stem revision in the MoM setting and a worst case scenario whereby lost to follow up were presumed to have failed. True stem failure was considered if revision occurred for a stem related complication.
1465 stems were included (1310 patients, 155 bilateral). The bearing surface was cobalt chrome on cobalt chrome in 1351 cases (92%). Seven hips were lost to follow up. Thirty-two stems (31 part of a MoM THR) underwent revision for any cause. Kaplan Meier survival analysis demonstrates an overall 97.4% survivorship. Subset analysis demonstrates 100% survivorship for aseptic loosening, 97.3% in the MoM setting and 96.7% for the worst case senario. Of the 32 cases of stem revision, only 13 were classified as 'true' stem failure.
This study represents the largest cohort of this uncemented femoral component with a minimum follow-up longer than 10 years. Our results demonstrate excellent long-term survivorship even in the presence of a challenging MoM environment.
我们评估了单中心 12.5 年(10.1-15.8 年)平均随访的近端羟基磷灰石涂层、双锥形、钛合金股骨柄的生存率。大多数柄作为金属对金属(MoM)全髋关节置换(THR)的一部分插入。
数据在本地数据库中进行前瞻性收集。对 2003 年至 2010 年间接受该假体初次 THR 的所有患者进行回顾性回顾。主要结果是任何原因导致的柄翻修。还分析了无菌性松动的柄翻修、MoM 环境中的柄翻修以及最糟糕的情况,即失访被认为是失败。如果发生与柄相关的并发症,则认为是真正的柄失败。
共纳入 1465 例(1310 例患者,155 例双侧)。1351 例(92%)的承载面为钴铬对钴铬。7 例髋关节失访。32 例(31 例为 MoM THR 的一部分)因任何原因接受了翻修。Kaplan-Meier 生存分析显示总体 97.4%的生存率。亚组分析显示无菌性松动的 100%生存率、MoM 环境下的 97.3%生存率和最差情况下的 96.7%生存率。在 32 例柄翻修中,只有 13 例被归类为“真正”的柄失败。
本研究代表了这种非骨水泥股骨组件的最大队列,随访时间超过 10 年。我们的结果表明,即使在具有挑战性的 MoM 环境下,也具有出色的长期生存率。