Wilson J A, Barratt A J, Gray J, Statham G B
Department of Microbiology, Central Pathology Laboratory, Stoke-on-Trent.
J Clin Pathol. 1988 Jun;41(6):679-82. doi: 10.1136/jcp.41.6.679.
A single bottle blood culture system (Oxoid Signal system) was compared with a conventional two bottle subculturing system. A total of 2016 routine blood samples yielded 186 (9%) clinically important isolates. Of these, 40 (21%) were isolated only in the Oxoid system and 30 (17%) only in the conventional system. One hundred and sixteen (62%) were isolated from both systems. Volume of blood was not significantly associated with the rate of detection but was significantly associated with the speed of detection. A continuation of the study with a Signal bottle and single conventional bottle confirmed initial results of the increased isolation rate using two systems. The Oxoid single bottle blood culture system was at least comparable with a conventional system in terms of rate of detection of organisms and was superior in terms of speed of detection.