与BACTEC放射测量系统相比,顶空气氛改变对Oxoid信号血培养系统的产量和检测速度的影响。

Effect of altered headspace atmosphere on yield and speed of detection of the Oxoid Signal blood culture system versus the BACTEC radiometric system.

作者信息

Weinstein M P, Mirrett S, Reimer L G, Reller L B

机构信息

Microbiology Laboratory, Robert Wood Johnson University Hospital, New Brunswick, New Jersey.

出版信息

J Clin Microbiol. 1990 Apr;28(4):795-7. doi: 10.1128/jcm.28.4.795-797.1990.

Abstract

The one-bottle Oxoid Signal blood culture system altered to provide a more aerobic bottle headspace was reassessed in a comparative study versus the two-bottle BACTEC radiometric system in 5,426 blood cultures. The BACTEC system detected more microorganisms (P less than 0.02), particularly anaerobes (P less than 0.05) and fungi (P less than 0.05).

摘要

一种改良后的Oxiold单瓶血液培养系统(其瓶头空间更利于需氧菌生长)在一项对比研究中与BACTEC双瓶放射测量系统进行了重新评估,该项研究共涉及5426份血培养样本。BACTEC系统检测到了更多的微生物(P<0.02),尤其是厌氧菌(P<0.05)和真菌(P<0.05)。