Lehti Leena, Söderberg Marcus, Mellander Helena, Wassélius Johan
Department of Clinical Sciences, Lund University, Lund, Sweden.
Vascular Center, Skåne University Hospital, Malmö, Sweden.
Eur J Radiol Open. 2020 Sep 4;7:100255. doi: 10.1016/j.ejro.2020.100255. eCollection 2020.
Onyx® embolization causes severe artifacts on subsequent CT-examinations, thereby seriously limiting the diagnostic quality.The purpose of this work was to compare the diagnostic quality of the tailored metal artifact reducing algorithms iMAR to standard reconstructions of CTA in patients treated with Onyx® embolization.
Twelve consecutive patients examined with Dual Energy CTA after Onyx® embolization were included. One standard image dataset without iMAR, and eight image datasets with different iMAR algorithms were reconstructed. Mean attenuation and noise were measured in the aorta or iliac arteries close to the Onyx® glue-cast and compared to the reference level in the diaphragmatic aorta. Mean attenuation and noise were also measured in the psoas muscle close to the Onyx®-glue and compared to the reference level in the psoas muscle at the level of the diaphragm.Subjective image quality and severity of artifacts was assessed by two experienced interventional radiologists blinded to reconstruction details.
All iMAR reconstructions had less distortion of the attenuation than the standard reconstructions and were also rated significantly better than the standard reconstructions by both interventional radiologists.
The iMAR algorithms can significantly reduce metal artifacts and improve the diagnostic quality in CTA in patients treated with Onyx® embolization, in many cases restoring non-diagnostic examinations to acceptable diagnostic quality.
Onyx®栓塞术会在后续CT检查中产生严重伪影,从而严重限制诊断质量。本研究的目的是比较在接受Onyx®栓塞术治疗的患者中,定制的金属伪影减少算法iMAR与CTA标准重建的诊断质量。
纳入12例连续接受Onyx®栓塞术后双能量CTA检查的患者。重建了一个未使用iMAR的标准图像数据集,以及八个使用不同iMAR算法的图像数据集。在靠近Onyx®胶水铸型的主动脉或髂动脉中测量平均衰减和噪声,并与膈主动脉中的参考水平进行比较。还在靠近Onyx®胶水的腰大肌中测量平均衰减和噪声,并与膈肌水平腰大肌中的参考水平进行比较。由两名对重建细节不知情的经验丰富的介入放射科医生评估主观图像质量和伪影严重程度。
所有iMAR重建的衰减失真均小于标准重建,且两名介入放射科医生对其的评分也显著高于标准重建。
iMAR算法可显著减少接受Onyx®栓塞术治疗患者CTA中的金属伪影并提高诊断质量,在许多情况下可将无法诊断的检查恢复到可接受的诊断质量。