Mathiesen E B, Lindgren J U, Reinholt F P, Sudmann E
J Biomed Mater Res. 1987 Apr;21(4):459-66. doi: 10.1002/jbm.820210406.
The tissue reaction to implanted materials and their wear products is important for the long-term fixation of a total hip replacement (THR). Recent reports indicate a relatively high incidence of aseptic loosening with Christiansen total hip prostheses. We examined tissue reactions to wear products in two groups of prostheses: 16 Christiansen prostheses with polyacetal (Delrin) sockets, and 18 Charnley-Müller total hip prostheses with ultrahigh-molecular-weight polyethylene (UHMWP) sockets. Specimens from these 34 patients were compared in a combined qualitative and semiquantitative morphological study. We saw similar histological changes in specimens of both groups, but specimens from the bone-cement interfaces of the acetabulum and femur shaft in the polyacetal group showed more inflammation and necrosis.
组织对植入材料及其磨损产物的反应对于全髋关节置换术(THR)的长期固定很重要。最近的报告表明,克里斯蒂安森全髋关节假体的无菌性松动发生率相对较高。我们研究了两组假体对磨损产物的组织反应:16个带有聚缩醛(Delrin)髋臼杯的克里斯蒂安森假体,以及18个带有超高分子量聚乙烯(UHMWP)髋臼杯的查恩利-米勒全髋关节假体。在一项定性和半定量形态学联合研究中,对这34例患者的标本进行了比较。我们在两组标本中观察到了相似的组织学变化,但聚缩醛组髋臼和股骨干骨水泥界面的标本显示出更多的炎症和坏死。