Maguire J K, Coscia M F, Lynch M H
Clin Orthop Relat Res. 1987 Mar(216):213-23.
Two cases of severe foreign body reaction to polymeric debris occurred following cemented total hip arthroplasty. The debris consisted of microfragments of both polyethylene and polymethylmethacrylate. Microfragments are believed to be primarily responsible for the extensive lytic reactions, which represent an extreme degree of the chronic inflammatory response that is commonly associated with loosening of cemented total hip arthroplasties. A schematic representation of the pathogenesis of component loosening, including the foreign body reaction, illustrates the relationship of this phenomenon to the general problem of aseptic loosening. Revision arthroplasty was successful in both cases, followed for a period of nine and 16 months, respectively.
在骨水泥型全髋关节置换术后发生了两例对聚合物碎片的严重异物反应。碎片由聚乙烯和聚甲基丙烯酸甲酯的微碎片组成。微碎片被认为是广泛溶骨反应的主要原因,溶骨反应代表了与骨水泥型全髋关节置换术松动通常相关的慢性炎症反应的极端程度。组件松动发病机制的示意图,包括异物反应,说明了这种现象与无菌性松动这一普遍问题的关系。两例翻修关节成形术均获成功,分别随访了9个月和16个月。