Suppr超能文献

[人体血浆精氨酸加压素的放射免疫测定。两种提取方法的比较]

[Radioimmunological assay of plasma arginine-vasopressin in man. Comparison of 2 methods of extraction].

作者信息

Wolf J P, Dumoulin G, Henriet M T, Baulay A, Berthelay S

出版信息

Pathol Biol (Paris). 1987 Apr;35(4):422-6.

PMID:3295707
Abstract

Two methods of extraction, prior to radio-immunoassay (RIA) of human plasma arginine vasopressin (AVP) were tested: ethanol extraction (ETH), chromatography with ODS-Silica (ODS-Sil). Recovery of 125I AVP was higher when using ODS-Sil than when using ETH. Recovery of standard AVP and plasma enriched with standard AVP was found to be more efficient and reproducible for ODS-Sil. However, the RIA used, performed after chromatography with ODS-Sil, is not enough sensitive to detect low concentrations but is able to detect high concentrations and physiological variations of plasma AVP.

摘要

在对人血浆精氨酸加压素(AVP)进行放射免疫分析(RIA)之前,测试了两种提取方法:乙醇提取(ETH)、ODS-硅胶柱色谱法(ODS-Sil)。使用ODS-Sil时,125I-AVP的回收率高于使用ETH时。对于ODS-Sil,标准AVP和富含标准AVP的血浆的回收率更高且可重复。然而,在使用ODS-Sil柱色谱法之后进行的RIA,检测低浓度时灵敏度不足,但能够检测血浆AVP的高浓度及生理变化。