Bistolfi Alessandro, Ferracini Riccardo, Aprato Alessandro, Massè Alessandro, Daghino Walter, Lea Sara, Artiaco Stefano, Lee Gwo-Chin
AO Città Della Salute e Della Scienza. Department of Orthopeadics, Traumatology and Rehabilitative Medicine, CTO Hospital, Turin, Italy.
Hoelloker Orthopaedics, Turin Italy. University of the Studies of Genova. Genova, Italy.
J Orthop. 2020 Aug 28;22:397-401. doi: 10.1016/j.jor.2020.08.023. eCollection 2020 Nov-Dec.
to evaluate the results of Delta ceramic-on-ceramic (CoC) for total-hip-arthroplasty (THA).
261 THA using Delta-CoC, retrospectively analyzed. A 36 mm head was used in 189 cases and a 32/40 mm in the others. The series have been compared to a group of 89 THA with Forte-CoC.
The Harris-Hip-Score improved from 49.1 ± 14.3 to 92.0 ± 8.9 (P < 0.001). In the Delta group there were one ceramic fracture and 2 dislocations. Two hips underwent revision. There were one revision in the Forte group for instability and one squeaking hip.
The new ceramic bearings provides a safe bearing for THA, with rare complications.
评估Delta陶瓷对陶瓷(CoC)全髋关节置换术(THA)的结果。
对261例使用Delta-CoC的THA进行回顾性分析。189例使用36毫米的股骨头,其余使用32/40毫米的股骨头。该系列与一组89例使用Forte-CoC的THA进行了比较。
Harris髋关节评分从49.1±14.3提高到92.0±8.9(P<0.001)。Delta组有1例陶瓷骨折和2例脱位。2例髋关节进行了翻修。Forte组有1例因不稳定进行翻修,1例髋关节有摩擦声。
新型陶瓷轴承为THA提供了安全的轴承,并发症罕见。