Hering Dominik Alexander, Kröger Kai, Bauer Ralf W, Eich Hans Theodor, Haverkamp Uwe
Department of Radiation Oncology, University Hospital of Muenster, Muenster, Germany.
RNS, Private Radiology and Radiation Therapy Group, Wiesbaden, Germany.
Br J Radiol. 2020 Dec 1;93(1116):20200152. doi: 10.1259/bjr.20200152. Epub 2020 Oct 1.
This work aims to investigate whether virtual non-contrast (VNC) dual-energy CT(DECT) of contrasted lung tumours can be used as an alternative for true non-contrast (TNC) images in radiotherapy. Two DECT techniques and a TNC CT were compared and influences on gross tumour volume (GTV) volume and CT number from motion artefacts in three-dimensional printed lung tumour models (LTM) in amotion phantom were examined.
Two spherical LTMs (diameter 3.0 cm) with different inner shapes were created in a three-dimensional printer. The inner shapes contained water or iodine (concentration 5 mg ml) and were scanned with a dual-source DECT (ds-DECT), single-source sequential DECT (ss-DECT) and TNC CT in a respiratory motion phantom (15 breaths/min, amplitude 1.5 cm). CT number and volume of LTMs were measured. Therefore, two GTVs were contoured.
Deviations in GTV volume (outer shape) of LTMs in motion for contrast-enhanced ss-DECT and ds-DECT VNC images compared to TNC images are not significant ( > 0.05). Relative GTV volume and CT number deviations (inner shapes) of LTMs in motion were 6.6 ± 0.6% and 104.4 ± 71.2 HU between ss-DECT and TNC CT and -8.4 ± 10.6% and 25.5 ± 58.5 HU between ds-DECT and TNC, respectively.
ss-DECT VNC images could not sufficiently subtract iodine from water in LTMs inmotion, whereas ds-DECT VNC images might be a valid alternative to a TNC CT.
ds-DECT provides a contrasted image for contouring and a non-contrasted image for radiotherapy treatment planning for LTM in motion.
本研究旨在探讨对比增强的肺部肿瘤虚拟非增强(VNC)双能CT(DECT)是否可作为放射治疗中真正非增强(TNC)图像的替代方案。比较了两种DECT技术和一种TNC CT,并研究了运动体模中三维打印肺肿瘤模型(LTM)中运动伪影对大体肿瘤体积(GTV)体积和CT值的影响。
在三维打印机中制作两个具有不同内部形状的球形LTM(直径3.0 cm)。内部形状包含水或碘(浓度5 mg/ml),并在呼吸运动体模(15次呼吸/分钟,幅度1.5 cm)中用双源DECT(ds-DECT)、单源序列DECT(ss-DECT)和TNC CT进行扫描。测量LTM的CT值和体积。因此,勾勒出两个GTV。
与TNC图像相比,对比增强的ss-DECT和ds-DECT VNC图像在运动中的LTM的GTV体积(外形)偏差不显著(>0.05)。运动中LTM的相对GTV体积和CT值偏差(内部形状)在ss-DECT和TNC CT之间分别为6.6±0.6%和104.4±71.2 HU,在ds-DECT和TNC之间分别为-8.4±10.6%和25.5±58.5 HU。
ss-DECT VNC图像在运动的LTM中不能充分从水中减去碘,而ds-DECT VNC图像可能是TNC CT的有效替代方案。
ds-DECT为运动中的LTM提供用于勾勒轮廓的对比增强图像和用于放射治疗计划的非增强图像。