Suppr超能文献

2298 例 Trifecta 主动脉瓣假体的耐久性和性能:倾向匹配分析。

Durability and Performance of 2298 Trifecta Aortic Valve Prostheses: A Propensity-Matched Analysis.

机构信息

Case Western Reserve University School of Medicine, Cleveland, Ohio.

Cleveland Clinic Lerner College of Medicine, Cleveland, Ohio.

出版信息

Ann Thorac Surg. 2021 Apr;111(4):1198-1205. doi: 10.1016/j.athoracsur.2020.07.040. Epub 2020 Oct 1.

Abstract

BACKGROUND

Reports of early failure of the Trifecta externally wrapped, bovine pericardial aortic valve prosthesis (Abbott Laboratories, Abbott Park, IL) raise concerns about its durability. This study evaluated the hemodynamic performance and explant of Trifecta valves compared with the PERIMOUNT bovine pericardial prosthesis (Edwards Lifesciences, Irvine, CA).

METHODS

From October 2007 to July 2017, 2305 patients received a Trifecta bioprosthesis during aortic valve replacement at Cleveland Clinic. Trends in postoperative valve hemodynamics were assessed from 4971 transthoracic echocardiograms and valve explants by systemic follow-up. To compare outcomes, 2298 patients receiving a Trifecta valve were 1:1 propensity matched from 17,281 patients receiving a PERIMOUNT bioprosthesis.

RESULTS

Mean age at implant was 69 years in both matched groups. Compared with PERIMOUNT valves, early transvalvular mean gradient of Trifecta valves was lower (11 vs 15 mm Hg at 1 year, P < .001); however, its longitudinal rate of rise was greater (P < .001), resulting in 5-year mean gradients of 17 vs 16 mm Hg, and more patients experienced severe aortic regurgitation (2.4% vs 0.81%; P < .001). At 5 years, 35 Trifecta valves had been explanted vs 14 PERIMOUNT valves; freedom from explant at 1, 3, and 5 years was 98.9%, 98.0%, and 95.9%, respectively, for the Trifecta group vs 99.3%, 99.0%, and 98.7% for the PERIMOUNT group (P < .001).

CONCLUSIONS

Compared with an older-generation internally mounted bovine pericardial valve, the Trifecta externally wrapped bioprosthesis exhibits superior early hemodynamic performance, but has a rapid increase in transvalvular gradient and more aortic regurgitation, with lower freedom from explant at 5 years. These findings raise concern regarding long-term Trifecta durability despite favorable early hemodynamics.

摘要

背景

有报道称,Trifecta 外部包裹牛心包主动脉瓣假体(雅培实验室,雅培公园,IL)早期失败,这引发了人们对其耐久性的担忧。本研究评估了 Trifecta 瓣膜与 PERIMOUNT 牛心包假体(爱德华兹生命科学公司,欧文,CA)的血流动力学性能和解剖结果。

方法

2007 年 10 月至 2017 年 7 月,克利夫兰诊所 2305 例患者在主动脉瓣置换术中接受了 Trifecta 生物瓣。通过系统随访,从 4971 次经胸超声心动图和瓣膜解剖中评估术后瓣膜血流动力学的趋势。为了比较结果,从 17281 例接受 PERIMOUNT 生物瓣的患者中,对接受 Trifecta 瓣膜的 2298 例患者进行了 1:1 的倾向评分匹配。

结果

两组患者的平均年龄均为 69 岁。与 PERIMOUNT 瓣膜相比,Trifecta 瓣膜的早期跨瓣平均梯度较低(1 年时为 11mmHg 比 15mmHg,P <.001);然而,其纵向上升率较高(P <.001),导致 5 年平均梯度为 17mmHg 比 16mmHg,更多患者出现严重主动脉瓣反流(2.4%比 0.81%;P <.001)。5 年内,有 35 个 Trifecta 瓣膜被解剖,而有 14 个 PERIMOUNT 瓣膜被解剖;Trifecta 组的 1、3、5 年无解剖生存率分别为 98.9%、98.0%和 95.9%,PERIMOUNT 组分别为 99.3%、99.0%和 98.7%(P <.001)。

结论

与新一代内置牛心包瓣膜相比,Trifecta 外部包裹生物瓣具有更好的早期血流动力学性能,但跨瓣梯度增加较快,主动脉瓣反流较多,5 年无解剖生存率较低。尽管早期血流动力学良好,但这些发现令人担忧 Trifecta 的长期耐久性。