Pulse Oximetry at the Wrist During Sleep: Performance, Challenges and Perspectives.

作者信息

Braun Fabian, Theurillat Patrick, Proenca Martin, Lemkaddem Alia, Ferrario Damien, De Jaegere Kurt, Horvath Christian M, Roth Corinne, Brill Anne-Kathrin, Lemay Mathieu, Ott Sebastian R

出版信息

Annu Int Conf IEEE Eng Med Biol Soc. 2020 Jul;2020:5115-5118. doi: 10.1109/EMBC44109.2020.9176081.

Abstract

The state-of-the-art non-invasive measurement of peripheral oxygen saturation (SpO2) during sleep is mainly based on pulse oximetry at the fingertip. Although this approach is noninvasive, it can still be obtrusive and cumbersome to apply, in particular for ambulatory monitoring over several nights.We developed a wrist-worn reflectance pulse oximetry device which can be embedded in a watch, making it less obtrusive and easy to apply. This device was tested in an ongoing clinical study on 57 subjects (33 patients and 24 healthy volunteers) undergoing a full overnight polysomnography recording. The accuracy was evaluated against state-of-the-art fingertip SpO2 measurements.In the 54 subjects available for analysis we obtained an SpO2 accuracy (ARMS) of 3.4 % when automatically rejecting 17.7 % of signals due to low quality. When further excluding measurements suffering from insufficient contact of the watch with the skin an ARMS of 2.7 % was obtained while rejecting a total of 23.2 % measurements. These accuracies comply with the ISO standard and the FDA guidance for pulse oximeters.The present results are promising and pave the way for unobtrusive and continuous monitoring of SpO2 to screen for sleep disordered breathing. Nonetheless, contact pressure and venous blood have shown to adversely affect the SpO2 estimation and remain a challenge for wrist-based reflectance pulse oximetry.

摘要

睡眠期间外周血氧饱和度(SpO2)的先进无创测量主要基于指尖脉搏血氧测定法。尽管这种方法是无创的,但应用起来仍然可能会造成干扰且不方便,特别是对于连续数晚的动态监测而言。我们开发了一种可嵌入手表的腕戴式反射式脉搏血氧测定设备,使其干扰性更小且易于应用。该设备正在一项针对57名受试者(33名患者和24名健康志愿者)的正在进行的临床研究中进行测试,这些受试者均接受了完整的夜间多导睡眠图记录。其准确性是对照最先进的指尖SpO2测量进行评估的。在可供分析的54名受试者中,由于信号质量低而自动剔除17.7%的信号时,我们获得的SpO2准确性(平均绝对相对误差,ARMS)为3.4%。当进一步排除手表与皮肤接触不充分的测量数据时,在总共剔除23.2%的测量数据的情况下,获得了2.7%的ARMS。这些准确性符合ISO标准以及FDA对脉搏血氧仪的指导方针。目前的结果很有前景,为无创且连续监测SpO2以筛查睡眠呼吸障碍铺平了道路。尽管如此,接触压力和静脉血已被证明会对SpO2估计产生不利影响,并且仍然是基于手腕的反射式脉搏血氧测定法面临的一项挑战。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验