Strong A A, Hazucha M J, Lundgren D A, Cerini E R
Med Instrum. 1987 Jun;21(3):189-94.
Most commercially available aerosol generators widely used in medical applications produce aerosols characterized by a large mass median diameter in the 4-8 micron range and the particle size in the 0.1-10.0 microns range. The desirable size of therapeutic and diagnostic aerosols, however, is about 2-4 microns mass median diameter, and less than 2.0 geometric standard deviation; this size increases the reproducibility of inhalation tests and enhances drug efficacy. We combined the commercially available DeVilbiss Model 65 nebulizer with a dilution/mixing chamber developed in our laboratory. The characteristics of this aerosol generator system were examined over a range of operating conditions and concentrations of solutions of three bronchoconstrictive agents--histamine, carbachol, and methacholine. The aerosol generator system produced a polydispersed aerosol with a mass median diameter range of 1.7-2.4 microns and geometric standard deviation of 1.5. The reliable and reproducible operation of the aerosol generator system greatly increases the power of bronchial challenge tests with bronchoconstrictive drugs.
大多数广泛应用于医学领域的市售气溶胶发生器所产生的气溶胶,其特点是质量中值直径大多在4 - 8微米范围内,粒径在0.1 - 10.0微米范围内。然而,治疗和诊断用气溶胶的理想尺寸是质量中值直径约为2 - 4微米,几何标准差小于2.0;这个尺寸能提高吸入试验的可重复性并增强药物疗效。我们将市售的德维比斯65型雾化器与我们实验室研发的稀释/混合室相结合。在一系列操作条件以及三种支气管收缩剂(组胺、卡巴胆碱和乙酰甲胆碱)溶液浓度下,对该气溶胶发生器系统的特性进行了研究。该气溶胶发生器系统产生了一种多分散气溶胶,其质量中值直径范围为1.7 - 2.4微米,几何标准差为1.5。该气溶胶发生器系统可靠且可重复的运行极大地提高了使用支气管收缩药物进行支气管激发试验的效能。