Abstract

A large amount of historical data regarding urinary iodine concentration (UIC) were measured with the Sandell-Kolthoff (S-K) method for iodine nutrition surveillance. The congruence in urinary iodine measurements between inductively coupled plasma mass spectrometry (ICP-MS) and the S-K method has been debated. A total of 2064 adult urine samples were included in the present study. The UIC measurement results obtained simultaneously by standardized ICP-MS and the S-K method were analyzed. The UIC obtained with ICP-MS was significantly higher than that obtained with the S-K method (158 μg/L vs. 148 μg/L, p < 0.001). The Bland-Altman difference plot showed a small but significant mean difference of 6.12 μg/L between the two methods. The stratified analysis showed that the correlation coefficient was higher in the UIC < 300 μg/L group than the UIC ≥ 300 μg/L group (0.93 vs. 0.88, p = 0.0001). The mean difference between the S-K and ICP-MS methods was positively correlated with the UIC. The ICP-MS and S-K methods were comparable when the UIC was less than 300 μg/L; however, UIC values between 300 and 600 μg/L should be compared with caution after considering the research objective. We do not suggest comparing UICs obtained from the ICP-MS and S-K methods in iodine monitoring studies if the UIC is greater than 600 μg/L.

摘要

大量的尿碘浓度(UIC)历史数据是用 Sandell-Kolthoff(S-K)法测量的,用于碘营养监测。电感耦合等离子体质谱法(ICP-MS)和 S-K 法在尿碘测量方面的一致性一直存在争议。本研究共纳入 2064 例成人尿样。同时分析了标准化 ICP-MS 和 S-K 法测量的 UIC 结果。ICP-MS 法测量的 UIC 明显高于 S-K 法(158μg/L 比 148μg/L,p<0.001)。Bland-Altman 差值图显示两种方法之间存在 6.12μg/L 的小但有统计学意义的平均差异。分层分析显示,在 UIC<300μg/L 组中,两种方法的相关系数高于 UIC≥300μg/L 组(0.93 比 0.88,p=0.0001)。S-K 法与 ICP-MS 法之间的平均差异与 UIC 呈正相关。当 UIC 小于 300μg/L 时,ICP-MS 和 S-K 法具有可比性;然而,在考虑研究目的后,对于 300-600μg/L 的 UIC 值,应谨慎比较。如果 UIC 大于 600μg/L,我们不建议在碘监测研究中比较 ICP-MS 和 S-K 法获得的 UIC 值。