In Vitro Evaluation of a Vibrating-Mesh Nebulizer Repeatedly Use over 28 Days.

作者信息

Lin Hui-Ling, Chen Chi-Shuo, Fink James B, Lee Guo-Hao, Huang Chun-Wei, Chen Jui-Chi, Chiang Zi Yi

机构信息

Department of Respiratory Therapy, Collage of Medicine, Chang Gung University, Taoyuan 33301, Taiwan.

Department of Respiratory Therapy, Chang Gung Memorial Hospital, Chiayi 61301, Taiwan.

出版信息

Pharmaceutics. 2020 Oct 15;12(10):971. doi: 10.3390/pharmaceutics12100971.

Abstract

This in vitro study evaluates the performance of a disposable vibrating-mesh nebulizer when used for 28 days. A lung model was used to simulate the breathing pattern of an adult with chronic obstructive pulmonary disease. The vibrating-mesh nebulizer was used for three treatments/day over 28 days without cleaning after each test. Results showed that the inhaled drug dose was similar during four weeks of use ( = 0.157), with 16.73 ± 4.46% at baseline and 15.29 ± 2.45%, 16.21 ± 2.21%, 17.56 ± 1.98%, and 17.13 ± 1.81%, after the first, second, third, and fourth weeks, respectively. The particle size distribution, residual drug volume, and nebulization time remained similar across four weeks of use ( = 0.110, = 0.763, and = 0.573, respectively). Mesh was inspected using optical microscopy and showed that approximately 50% of mesh pores were obscured after 84 runs, and light penetration through the aperture plate was significantly reduced after the 21st use ( < 0.001) with no correlation to nebulizer performance. We conclude that the vibrating-mesh nebulizer delivered doses of salbutamol solution effectively over four weeks without cleaning after each use even though the patency and clarity of the aperture plate were reduced by the first week of use.

摘要

这项体外研究评估了一次性振动网孔雾化器使用28天时的性能。使用肺部模型模拟慢性阻塞性肺疾病成人的呼吸模式。振动网孔雾化器在28天内每天用于三次治疗,每次测试后不进行清洁。结果显示,在使用四周期间吸入药物剂量相似(=0.157),基线时为16.73±4.46%,在第一、第二、第三和第四周后分别为15.29±2.45%、16.21±2.21%、17.56±1.98%和17.13±1.81%。在使用四周期间,粒径分布、残留药物体积和雾化时间保持相似(分别为=0.110、=0.763和=0.573)。使用光学显微镜检查网孔,结果显示在84次运行后约50%的网孔被堵塞,在第21次使用后通过孔板的光穿透率显著降低(<0.001),且与雾化器性能无关。我们得出结论,尽管孔板的通畅性和清晰度在使用第一周后有所降低,但振动网孔雾化器在每次使用后不进行清洁的情况下,仍能在四周内有效输送沙丁胺醇溶液剂量。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/1307/7602390/b79a24b7e4f8/pharmaceutics-12-00971-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验