Department of Anesthesiology, Erasmus MC, University Medical Center Rotterdam, Rotterdam, The Netherlands.
Department of Pediatric Surgery, Erasmus MC-Sophia Children's Hospital, University Medical Center Rotterdam, Rotterdam, The Netherlands.
J Clin Monit Comput. 2021 Dec;35(6):1357-1366. doi: 10.1007/s10877-020-00602-y. Epub 2020 Oct 21.
Mitochondria are the primary consumers of oxygen and therefore an important location for oxygen availability and consumption measurement. A technique has been developed for mitochondrial oxygen tension (mitoPO) measurement, incorporated in the COMET. In contrast to most textbooks, relatively high average mitoPO values have been reported. The first aim of this study was to verify the validity of the COMET calibration for mitoPO measurements in human skin. The second aim was to compare the dynamics of mitoPO to several other techniques assessing tissue oxygenation. Firstly, we performed a two-point calibration. Mitochondrial oxygen depletion was achieved with vascular occlusion. A high mitoPO was reached by local application of cyanide. MitoPO was compared to the arterial oxygen partial pressure (PaO). Secondly, for deoxygenation kinetics we compared COMET variables with the LEA O2C, SenTec OxiVenT™ and Medtronic INVOS™ parameters during a vascular occlusion test. 20 healthy volunteers were recruited and resulted in 18 datasets (2 times 9 subjects). The lowest measured mitoPO value per subject had a median [IQR] of 3.0 [1.0-4.0] mmHg, n = 9. After cyanide application the mitoPO was 94.1 mmHg [87.2-110.9] and did not differ significantly (n = 9, p = 0.5) from the PaO of 101.0 [98.0-106.0] mmHg. In contrast to O2C, OxiVenT™ and INVOS parameters, mitoPO declined within seconds with pressure on the probe. The kinetics from this decline are used to mitochondrial oxygen consumption (mitoVO). This study validates the calibration of the COMET device in humans. For mitoVO measurements not only blood flow cessation but application of local pressure is of great importance to clear the measurement site of oxygen-carrying erythrocytes.
线粒体是氧气的主要消耗者,因此是测量氧气供应和消耗的重要位置。已经开发出一种用于测量线粒体氧张力(mitoPO)的技术,该技术已集成到 COMET 中。与大多数教科书不同,已经报道了相对较高的平均 mitoPO 值。本研究的第一个目的是验证 COMET 校准在人体皮肤中的 mitoPO 测量中的有效性。第二个目的是比较 mitoPO 与评估组织氧合的其他几种技术的动力学。首先,我们进行了两点校准。通过血管闭塞实现线粒体氧耗竭。局部应用氰化物可达到高 mitoPO。将 mitoPO 与动脉氧分压(PaO)进行比较。其次,对于去氧动力学,我们在血管闭塞试验中比较了 COMET 变量与 LEA O2C、SenTec OxiVenT™和 Medtronic INVOS™参数。招募了 20 名健康志愿者,共获得 18 组数据集(2 次 9 名受试者)。每位受试者的最低测量 mitoPO 值中位数[IQR]为 3.0[1.0-4.0]mmHg,n=9。应用氰化物后,mitoPO 为 94.1mmHg[87.2-110.9],与 PaO 101.0mmHg[98.0-106.0]无显著差异(n=9,p=0.5)。与 O2C、OxiVenT™和 INVOS 参数相反,mitoPO 在探头受压后几秒钟内下降。从这种下降中得出的动力学用于测量线粒体耗氧量(mitoVO)。本研究验证了 COMET 设备在人体中的校准。对于 mitoVO 测量,不仅需要停止血流,而且局部压力的应用对于清除携氧红细胞的测量部位非常重要。