Department of Anesthesiology, Laboratory of Experimental Anesthesiology, Erasmus MC, University Medical Center Rotterdam, Rotterdam, The Netherlands.
PLoS One. 2023 Feb 9;18(2):e0278561. doi: 10.1371/journal.pone.0278561. eCollection 2023.
The newly introduced Cellular Oxygen METabolism (COMET®) monitor enables the measurement of mitochondrial oxygen tension (mitoPO2) using the protoporphyrin IX triplet state lifetime technique (PpIX-TSLT). This study aims to investigate the feasibility and applicability of the COMET® measurements in the operating theatre and study the behavior of the new parameter mitoPO2 during stable operating conditions.
In this observational study mitochondrial oxygenation was measured in 20 patients during neurosurgical procedures using the COMET® device. Tissue oxygenation and local blood flow were measured by the Oxygen to See (O2C). Primary outcomes included mitoPO2, skin temperature, mean arterial blood pressure, local blood flow and tissue oxygenation.
All patients remained hemodynamically stable during surgery. Mean baseline mitoPO2 was 60 ± 19 mmHg (mean ± SD) and mean mitoPO2 remained between 40-60 mmHg during surgery, but tended to decrease over time in line with increasing skin temperature.
This study presents the feasibility of mitochondrial oxygenation measurements as measured by the COMET® monitor in the operating theatre and shows the parameter mitoPO2 to behave in a stable and predictable way in the absence of notable hemodynamic alterations. The results provide a solid base for further research into the added value of mitochondrial oxygenation measurements in the perioperative trajectory.
新引入的细胞氧代谢(COMET®)监测仪可使用原卟啉 IX 三重态寿命技术(PpIX-TSLT)测量线粒体氧张力(mitoPO2)。本研究旨在探讨 COMET®测量在手术室中的可行性和适用性,并研究新参数 mitoPO2 在稳定手术条件下的行为。
在这项观察性研究中,使用 COMET®设备在 20 名神经外科手术患者中测量线粒体氧合作用。通过 Oxygen to See(O2C)测量组织氧合和局部血流。主要结局包括 mitoPO2、皮肤温度、平均动脉血压、局部血流和组织氧合。
所有患者在手术过程中血流动力学均保持稳定。平均基线 mitoPO2 为 60 ± 19 mmHg(平均值 ± 标准差),手术期间 mitoPO2 平均值保持在 40-60 mmHg 之间,但随着皮肤温度的升高, tended 呈下降趋势。
本研究介绍了 COMET®监测仪在手术室中测量线粒体氧合作用的可行性,并表明在没有明显血流动力学改变的情况下,参数 mitoPO2 的行为稳定且可预测。研究结果为进一步研究围手术期轨迹中线粒体氧合作用测量的附加值提供了坚实的基础。