通过Eclipse治疗计划系统评估前列腺癌患者滑动窗口调强放射治疗中的射野外剂量计算准确性。

Assessing the out-of-field dose calculation accuracy by eclipse treatment planning system in sliding window IMRT of prostate cancer patients.

作者信息

Faghihi Moghaddam Fahimeh, Bakhshandeh Mohsen, Ghorbani Mahdi, Mofid Bahram

机构信息

Biomedical Engineering and Medical Physics Department, Faculty of Medical Sciences, Shahid Beheshti University of Medical Sciences, Tehran, Iran.

Department of Radiation Technology, Faculty of Allied Medical Sciences, Shahid Beheshti University of Medical Sciences, Tehran, Iran.

出版信息

Comput Biol Med. 2020 Dec;127:104052. doi: 10.1016/j.compbiomed.2020.104052. Epub 2020 Oct 21.

Abstract

AIM

The objective of this study was to evaluate out-of-field dose distribution calculation accuracy by the Anisotropic Analytical Algorithm (AAA), version 13.0.26, in Eclipse TPS, (Varian Medical Systems, Palo Alto, Ca, USA) for sliding window IMRT delivery technique in prostate cancer patients.

MATERIALS AND METHODS

Prostate IMRT plans with nine coplanar were calculated with the AAA Eclipse treatment planning system. To assess the accuracy of dose calculation predicted by the Eclipse in normal tissue and OARs located out of radiation field areas, including the rectum, bladder, right and left head of the femur, absolute organ dose value, and dose distribution were measured using the Delta IMRT phantom.

RESULTS

In the out-of-field areas, underestimation of -0.66% in organs near the field edge to -39.63% in organs far from the field edge (2.5 and 7.3 cm respectively) occurred in the TPS calculations. The percentage of dose deviation for the femoral heads was 95.7 on average while for the organ closer to the target (rectum) it was 79.81.

CONCLUSIONS

AAA dosimetry algorithm (used in Eclipse TPS) showed poor dose calculation in areas beyond the treatment fields border where underestimation varies with the distance from the field edges. A significant underestimation was found for the AAA algorithm in the sliding window IMRT technique (P-value > 0.05).

摘要

目的

本研究的目的是评估美国瓦里安医疗系统公司位于加利福尼亚州帕洛阿尔托市的Eclipse治疗计划系统(TPS)中13.0.26版本的各向异性分析算法(AAA)对前列腺癌患者滑动窗口调强放疗(IMRT)技术的野外剂量分布计算准确性。

材料与方法

使用AAA Eclipse治疗计划系统计算九个共面的前列腺IMRT计划。为了评估Eclipse预测的剂量计算在位于辐射野外区域的正常组织和危及器官(OARs)中的准确性,包括直肠、膀胱、左右股骨头,使用Delta IMRT体模测量绝对器官剂量值和剂量分布。

结果

在野外区域,TPS计算中,靠近野外边缘的器官剂量低估了-0.66%,远离野外边缘的器官(分别为2.5厘米和7.3厘米)剂量低估了-39.63%。股骨头的剂量偏差百分比平均为95.7,而更靠近靶区的器官(直肠)为79.81。

结论

(Eclipse TPS中使用的)AAA剂量测定算法在治疗野外边界以外的区域显示出较差的剂量计算,低估程度随与野外边缘的距离而变化。在滑动窗口IMRT技术中发现AAA算法存在显著低估(P值>0.05)。