Molecularly imprinted polymer-based device for field collection of oral fluid samples for cocaine identification.

机构信息

Department of Analytical Chemistry, University of Valencia, 50th Dr. Moliner St., 46100 Burjassot, Spain.

Department of Analytical Chemistry, University of Valencia, 50th Dr. Moliner St., 46100 Burjassot, Spain.

出版信息

J Chromatogr A. 2020 Dec 6;1633:461629. doi: 10.1016/j.chroma.2020.461629. Epub 2020 Oct 17.

Abstract

In this paper, a low-cost, rapid, easy, and potentially portable tool for the identification of cocaine and its semi-quantitative determination in oral fluid has been proposed. A field collection device has been designed, based on a cotton pad with an indicator and a molecularly imprinted polymer (MIP) sorbent, to selective retain cocaine from oral fluid components. After sample collection, cocaine is transferred by using phosphate buffer to the MIP and then eluted with 2-propanol. The obtained extract is analysed by ion mobility spectrometry (IMS), providing a cut-off value of 20 µg L that identifies 100 % true-positive and 95 % true-negative samples. The MIP-IMS procedure has been validated by the analysis of oral fluid samples, collected from cocaine users at recreation environments, by comparing the results with lateral flow immunoassay and chromatographic reference methods. Thus, the proposed methodology allows a simple and fast cocaine identification that can be carried out in field by non-specialized personnel, such as health personnel, law enforcement bodies, and customs staff.

摘要

本文提出了一种低成本、快速、简便、具有潜在便携性的可卡因识别工具,并可对半定量测定口腔液中的可卡因。设计了一种基于带有指示剂和分子印迹聚合物(MIP)吸附剂的棉垫的现场采集装置,用于从口腔液成分中选择性保留可卡因。采集样品后,使用磷酸盐缓冲液将可卡因转移到 MIP 上,然后用异丙醇洗脱。所得提取物通过离子迁移谱(IMS)进行分析,给出了 20μg/L 的截止值,可识别 100%的真阳性和 95%的真阴性样本。通过比较侧向流动免疫测定和色谱参考方法分析在娱乐环境中采集的可卡因使用者的口腔液样本,对 MIP-IMS 程序进行了验证。因此,所提出的方法允许进行简单快速的可卡因识别,并且可以由非专业人员(如医务人员、执法机构和海关人员)在现场进行。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验