使用两种不同的谱域光学相干断层扫描血管造影设备测量健康受试者的浅表和深部黄斑无血管区面积

Superficial and Deep Foveal Avascular Zone Area Measurement in Healthy Subjects Using Two Different Spectral Domain Optical Coherence Tomography Angiography Devices.

作者信息

Anvari Pasha, Najafi Amin, Mirshahi Reza, Sardarinia Mahsa, Ashrafkhorasani Maryam, Kazemi Pegah, Aghai Gholamhoseyn, Habibi Abbas, Falavarjani Khalil Ghasemi

机构信息

Eye Research Center, The Five Senses Institute, Rassoul Akram Hospital, Iran University of Medical Sciences, Tehran, Iran.

出版信息

J Ophthalmic Vis Res. 2020 Oct 25;15(4):517-523. doi: 10.18502/jovr.v15i4.7791. eCollection 2020 Oct-Dec.

Abstract

PURPOSE

To compare the area of the foveal avascular zone (FAZ) in the superficial and deep retinal layers using two different spectral-domain optical coherence tomography angiography (OCTA) devices.

METHODS

A cross-sectional comparative study was conducted to obtain macular OCTA images from healthy subjects using Optovue RTVue XR Avanti (Optovue, Inc, Fremont, CA) and Spectralis HRA+OCTA (Heidelberg Engineering, Heidelberg, Germany). Two independent trained graders measured the FAZ area using automated slab segmentation. The FAZ area in the superficial and deep retinal layers were compared.

RESULTS

Twenty-three eyes of 23 subjects were included. The graders agreement was excellent ( 0.86) for all measurements. The mean FAZ area was significantly larger at the superficial retinal layer as compared to the deep retinal layer on both devices (0.31 0.08 mm vs 0.26 0.08 mm in Optovue and 0.55 0.16 mm vs 0.36 0.13 mm in Spectralis, both 0.001). The mean FAZ area was significantly greater in the superficial and deep retinal layers using Spectralis as compared to Optovue measurements ( 0.001 for both comparisons).

CONCLUSION

In contrast to previous reports, the FAZ area was larger in the superficial retina as compared to deep retinal layers using updated software versions. Measurements from different devices cannot be used interchangeably.

摘要

目的

使用两种不同的光谱域光学相干断层扫描血管造影(OCTA)设备比较视网膜浅层和深层中央凹无血管区(FAZ)的面积。

方法

进行一项横断面比较研究,使用Optovue RTVue XR Avanti(美国加利福尼亚州弗里蒙特市Optovue公司)和Spectralis HRA + OCTA(德国海德堡海德堡工程公司)从健康受试者获取黄斑OCTA图像。两名独立训练的分级人员使用自动分层分割测量FAZ面积。比较视网膜浅层和深层的FAZ面积。

结果

纳入23名受试者的23只眼。所有测量中分级人员的一致性都很好(0.86)。两种设备上,视网膜浅层的平均FAZ面积均显著大于深层(Optovue中分别为0.31±0.08mm²和0.26±0.08mm²,Spectralis中分别为0.55±0.16mm²和0.36±0.13mm²,均P<0.001)。与Optovue测量相比,使用Spectralis测量的视网膜浅层和深层的平均FAZ面积均显著更大(两次比较均P<0.001)。

结论

与之前的报告相反,使用更新软件版本时,视网膜浅层的FAZ面积大于深层。不同设备的测量结果不能互换使用。