Suppr超能文献

用于调强放射治疗的补偿滤过器下克拉克森法计算吸收剂量的优化

Optimization of Clarkson's Method for Calculating Absorbed Dose under Compensator Filters used in Intensity-modulated Radiation Therapy.

作者信息

M Pourkaveh, A Haghparast, M T Eivazi, K Ghazikhanlu Sani

机构信息

MSc, Medical Physics Department, Faculty of Medicine, Kermanshah University of Medical Sciences, Kermanshah, Iran.

PhD, Medical Physics Department, Faculty of Medicine, Kermanshah University of Medical Sciences, Kermanshah, Iran.

出版信息

J Biomed Phys Eng. 2020 Oct 1;10(5):575-582. doi: 10.31661/jbpe.v0i0.858. eCollection 2020 Oct.

Abstract

BACKGROUND

Intensity Modulated Radiation Therapy (IMRT) is extensively used in the treatment of malignancies. Clarkson's method is one of the leading methods for dose calculation at open points present in irregular fields.

OBJECTIVE

The aim of this study is to generalize the Clarkson's method for dose calculation at points under compensator filters in IMRT method and its application in IMRT quality control as well.

MATERIAL AND METHODS

In this experimental study, compensator filters were designed in two forms: flat filter and block piled-up compensator. The measurements for the compensator filters and open fields in 5 and10 cm depths at energy levels (6, 10 and 18 MV) and in fields with different dimensions were performed using "Mapcheck2" dosimeter. The aim of performing calculations is to derive the theoretical dose by the generalized Clarkson's equation and comparing it with data resulted from the measurement for confirming the Clarkson's equation presented.

RESULTS

These results demonstrate the data derived from the generalized Clarkson's method are in good agreement with the data resulted from measurement; the highest error of the proposed equation was 3% for flat filter, and less than 5% for block-piled-up filter. Higher error in the block-piled-up filter compared with the flat filter was due to the presence of leakage between these blocks.

CONCLUSION

The results of this study demonstrated that the presented equation is reliable and valid, and the proposed equation can be applied for dose calculation at all points under the compensator filter or the shielded areas.

摘要

背景

调强放射治疗(IMRT)广泛应用于恶性肿瘤的治疗。克拉克森方法是不规则野中开放点剂量计算的主要方法之一。

目的

本研究旨在推广克拉克森方法,用于调强放射治疗中补偿滤过器下各点的剂量计算及其在调强放射治疗质量控制中的应用。

材料与方法

在本实验研究中,补偿滤过器设计为两种形式:平板滤过器和块状堆积补偿器。使用“Mapcheck2”剂量仪对能量水平为(6、10和18兆伏)、深度为5厘米和10厘米、不同尺寸野的补偿滤过器和开放野进行测量。进行计算的目的是通过广义克拉克森方程得出理论剂量,并将其与测量数据进行比较,以确认所提出的克拉克森方程。

结果

这些结果表明,广义克拉克森方法得出的数据与测量数据高度吻合;所提出方程的最高误差,平板滤过器为3%,块状堆积滤过器小于5%。块状堆积滤过器的误差高于平板滤过器,这是由于这些块之间存在泄漏。

结论

本研究结果表明,所提出的方程可靠且有效,该方程可应用于补偿滤过器或屏蔽区域下所有点的剂量计算。