Suppr超能文献

Impella 5.5 与 Centrimag:设备血液相容性的头对头比较。

Impella 5.5 Versus Centrimag: A Head-to-Head Comparison of Device Hemocompatibility.

机构信息

From the Department of Medicine, Sarver Heart Center, University of Arizona, Tucson, Arizona.

Department of Mechanical Engineering, Vanderbilt University, Nashville, Tennessee.

出版信息

ASAIO J. 2020 Nov/Dec;66(10):1142-1151. doi: 10.1097/MAT.0000000000001283.

Abstract

Despite growing use of mechanical circulatory support, limitations remain related to hemocompatibility. Here, we performed a head-to-head comparison of the hemocompatibility of a centrifugal cardiac assist system-the Centrimag, with that of the latest generation of an intravascular microaxial system-the Impella 5.5. Specifically, hemolysis, platelet activation, microparticle (MP) generation, and von Willebrand factor (vWF) degradation were evaluated for both devices. Freshly obtained porcine blood was recirculated within device propelled mock loops for 4 hours, and alteration of the hemocompatibility parameters was monitored over time. We found that the Impella 5.5 and Centrimag exhibited low levels of hemolysis, as indicated by minor increase in plasma free hemoglobin. Both devices did not induce platelet degranulation, as no alteration of β-thromboglobulin and P-selectin in plasma occurred, rather minor downregulation of platelet surface P-selectin was detected. Furthermore, blood exposure to shear stress via both Centrimag and Impella 5.5 resulted in a minor decrease of platelet count with associated ejection of procoagulant MPs, and a decrease of vWF functional activity (but not plasma level of vWF-antigen). Greater MP generation was observed with the Centrimag relative to the Impella 5.5. Thus, the Impella 5.5 despite having a lower profile and higher impeller rotational speed demonstrated good and equivalent hemocompatibility, in comparison with the predicate Centrimag, with the advantage of lower generation of MPs.

摘要

尽管机械循环支持的应用日益广泛,但与血液相容性相关的局限性仍然存在。在这里,我们对离心式心脏辅助系统-Centrimag 和最新一代血管内微轴流系统-Impella 5.5 的血液相容性进行了头对头比较。具体来说,评估了这两种设备的溶血、血小板活化、微粒 (MP) 生成和血管性血友病因子 (vWF) 降解。新鲜获得的猪血液在设备推动的模拟回路中循环 4 小时,并监测随时间变化的血液相容性参数的变化。我们发现,Impella 5.5 和 Centrimag 的溶血水平较低,血浆游离血红蛋白仅略有增加。两种设备均未诱导血小板脱颗粒,因为血浆中β-血栓球蛋白和 P-选择素没有改变,而血小板表面 P-选择素的下调幅度较小。此外,血液通过 Centrimag 和 Impella 5.5 暴露于剪切应力导致血小板计数略有下降,同时伴有促凝 MP 的喷射,以及 vWF 功能活性下降(但血浆 vWF 抗原水平不变)。与 Impella 5.5 相比,Centrimag 产生的 MP 更多。因此,与前药 Centrimag 相比,尽管 Impella 5.5 具有更低的轮廓和更高的叶轮转速,但表现出良好且相当的血液相容性,其优点是生成的 MPs 更少。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/a7b6/7594535/9f80137e2f5a/mat-66-1142-g001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验