Mechanical Evaluation of the Stability of One or Two Miniscrews under Loading on Synthetic Bone.

作者信息

Pradal Andrea, Nucci Ludovica, Derton Nicola, De Felice Maria Elena, Turco Gianluca, Grassia Vincenzo, Contardo Luca

机构信息

Department of Medical, Surgical and Health Sciences, University of Trieste, 34127 Trieste, Italy.

Multidisciplinary Department of Medical-Surgical and Dental Specialties, University of Campania Luigi Vanvitelli, 81100 Naples, Italy.

出版信息

J Funct Biomater. 2020 Nov 5;11(4):80. doi: 10.3390/jfb11040080.

Abstract

The aim of the present study was to evaluate the primary stability of a two-miniscrew system inserted into a synthetic bone and to compare the system with the traditional one. Forty-five bi-layered polyurethane blocks were used to simulate maxillary cancellous and cortical bone densities. Samples were randomly assigned to three groups-one-miniscrew system (Group A, = 23), two-miniscrew system (Group B, = 22) and archwire-only (Group C, = 10). A total of 67 new miniscrews were subdivided into Group A (23 singles) and Group B (22 couples). 30 mm of 19″ × 25″ archwires were tied to the miniscrew. The load was applied perpendicularly to the archwire. Maximum Load Value (MLV), Yield Load (YL) and Loosening Load (LL) were recorded for each group. The YL of Group B and C had a mean value respectively of 4.189 ± 0.390 N and 3.652 ± 0.064 N. The MLV of Group A, B and C had a mean value respectively of 1.871 ± 0.318N, of 4.843 ± 0.515 N and 4.150 ± 0.086 N. The LL of Group A and B had a mean value respectively of 1.871 ± 0.318 N and of 2.294 ± 0.333 N. A two- temporary anchorage device (TAD) system is on average stiffer than a one-TAD system under orthodontic loading.

摘要

本研究的目的是评估植入合成骨的双微型螺钉系统的初始稳定性,并将该系统与传统系统进行比较。使用45个双层聚氨酯块模拟上颌松质骨和皮质骨密度。样本被随机分为三组:单微型螺钉系统(A组,n = 23)、双微型螺钉系统(B组,n = 22)和仅使用弓丝组(C组,n = 10)。总共67个新的微型螺钉被细分为A组(23个单颗)和B组(22对)。将30 mm的19″×25″弓丝结扎到微型螺钉上。载荷垂直施加于弓丝。记录每组的最大载荷值(MLV)、屈服载荷(YL)和松动载荷(LL)。B组和C组的YL平均值分别为4.189±0.390 N和3.652±0.064 N。A组、B组和C组的MLV平均值分别为1.871±0.318 N、4.843±0.515 N和4.150±0.086 N。A组和B组的LL平均值分别为1.871±0.318 N和2.294±0.333 N。在正畸加载下,双临时锚固装置(TAD)系统平均比单TAD系统更硬。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/8814/7711976/195d330e53f7/jfb-11-00080-g001.jpg