Investigation of safety for electrochemotherapy and irreversible electroporation ablation therapies in patients with cardiac pacemakers.

机构信息

University of Ljubljana, Faculty of Electrical Engineering, Trzaska 25, 1000, Ljubljana, Slovenia.

University of Zagreb, Faculty of Electrical Engineering and Computing, Unska 3, 10000, Zagreb, Croatia.

出版信息

Biomed Eng Online. 2020 Nov 16;19(1):85. doi: 10.1186/s12938-020-00827-7.

Abstract

BACKGROUND

The effectiveness of electrochemotherapy of tumors (ECT) and of irreversible electroporation ablation (IRE) depends on different mechanisms and delivery protocols. Both therapies exploit the phenomenon of electroporation of the cell membrane achieved by the exposure of the cells to a series of high-voltage electric pulses. Electroporation can be fine-tuned to be either reversible or irreversible, causing the cells to either survive the exposure (in ECT) or not (in IRE), respectively. For treatment of tissues located close to the heart (e.g., in the liver), the safety of electroporation-based therapies is ensured by synchronizing the electric pulses with the electrocardiogram. However, the use of ECT and IRE remains contraindicated for patients with implanted cardiac pacemakers if the treated tissues are located close to the heart or the pacemaker. In this study, two questions are addressed: can the electroporation pulses interfere with the pacemaker; and, can the metallic housing of the pacemaker modify the distribution of electric field in the tissue sufficiently to affect the effectiveness and safety of the therapy?

RESULTS

The electroporation pulses induced significant changes in the pacemaker ventricular pacing pulse only for the electroporation pulses delivered during the pacing pulse itself. No residual effects were observed on the pacing pulses following the electroporation pulses for all tested experimental conditions. The results of numerical modeling indicate that the presence of metal-encased pacemaker in immediate vicinity of the treatment zone should not impair the intended effectiveness of ECT or IRE even when the casing is in direct contact with one of the active electrodes. Nevertheless, the contact between the casing and the active electrode should be avoided due to significant tissue heating at the site of the other active electrode for the IRE protocol and may cause the pulse generator to fail to deliver the pulses due to excessive current draw.

CONCLUSIONS

The observed effects of electroporation pulses delivered in close vicinity of the pacemaker or its electrodes do not indicate adverse consequences for either the function of the pacemaker or the treatment outcome. These findings should contribute to making electroporation-based treatments accessible also to patients with implanted cardiac pacemakers.

摘要

背景

肿瘤电化学疗法(ECT)和不可逆电穿孔消融(IRE)的疗效取决于不同的机制和输送方案。这两种疗法都利用细胞膜电穿孔的现象,通过将细胞暴露于一系列高压电脉冲来实现。电穿孔可以微调为可逆或不可逆,分别导致细胞存活(在 ECT 中)或不存活(在 IRE 中)。对于靠近心脏的组织(例如肝脏)的治疗,通过将电脉冲与心电图同步,可以确保基于电穿孔的疗法的安全性。然而,如果治疗组织靠近心脏或起搏器,植入式心脏起搏器的患者仍然禁止使用 ECT 和 IRE。在这项研究中,提出了两个问题:电穿孔脉冲是否会干扰起搏器;以及起搏器的金属外壳是否会足以改变组织中电场的分布,从而影响治疗的有效性和安全性?

结果

只有在起搏脉冲本身期间输送的电穿孔脉冲才会导致电穿孔脉冲对起搏器心室起搏脉冲产生显著变化。在所有测试的实验条件下,在电穿孔脉冲之后,没有观察到对起搏脉冲的残留影响。数值建模的结果表明,即使起搏器外壳与其中一个活动电极直接接触,位于治疗区域附近的金属外壳起搏器的存在也不应损害 ECT 或 IRE 的预期有效性。然而,由于 IRE 方案中在另一个活动电极的部位存在显著的组织加热,应避免外壳与活动电极之间的接触,这可能会导致脉冲发生器因电流过大而无法输送脉冲。

结论

在起搏器或其电极附近输送的电穿孔脉冲的观察到的影响并未表明对起搏器的功能或治疗结果产生不利后果。这些发现应该有助于使基于电穿孔的治疗也可用于植入式心脏起搏器的患者。