Universiti Sultan Zainal Abidin, Faculty of Medicine, Kuala Terengganu, Terengganu, Malaysia.
Med J Malaysia. 2020 Nov;75(6):685-690.
To design and develop a simple vision test algorithm for mobile application and perform a pilot study to determine its validity and reliability as a tool for vision test in the community.
A simple visual acuity test algorithm in the form of a single letter E display was designed as the optotype for development of a mobile application. The standardised optotype is presented at random to test visual acuity for corresponding level of 3/60, 6/60, 6/18, and 6/12. The final result is auto-generated based on the classification of the WHO for visual impairment and blindness. The Snellen chart was used as the gold standard to determine its validity while five different users were involved to determine its inter-rater reliability. A pilot study was performed between April till November 2019, in the Universiti Sultan Zainal Abidin Medical Centre (UMC) at Kuala Nerus and Mooris Optometrist Centre at Marang, Terengganu. A total of 279 participants aged four years old and above were involved in this study.
The highest sensitivity was found at the vision level cut-off point of 6/12 with the percentage of 92.7% and 86.8% for the right and left eye, respectively. The specificity was more than 89% for all vision levels in both eyes. The Krippendorff's alpha value for the inter-rater reliability was 0.87 and 0.83.
The relatively high level of validity and reliability obtained indicate the feasibility of using the designed optotype to develop a valid and reliable mobile app for vision test. The app can be used to screen vision by nonmedical persons, at anytime and anywhere to help improve public awareness and capability to correctly determine their visual status.
设计并开发一种简单的视力测试算法,用于移动应用程序,并进行初步研究以确定其作为社区视力测试工具的有效性和可靠性。
设计了一种简单的视力测试算法,以单个字母 E 的形式显示,作为开发移动应用程序的视标。标准视标随机呈现,以测试相应 3/60、6/60、6/18 和 6/12 视力的视力。最终结果基于世界卫生组织对视力障碍和失明的分类自动生成。使用 Snellen 图表作为确定其有效性的金标准,而涉及五位不同的用户以确定其评分者间可靠性。2019 年 4 月至 11 月,在关丹的Universiti Sultan Zainal Abidin 医疗中心(UMC)和登嘉楼的 Mooris 验光中心进行了一项初步研究。共有 279 名年龄在四岁及以上的参与者参与了这项研究。
在视力水平截止点为 6/12 时,右眼和左眼的敏感性最高,分别为 92.7%和 86.8%。右眼和左眼的所有视力水平的特异性均超过 89%。评分者间可靠性的 Krippendorff's alpha 值为 0.87 和 0.83。
获得的相对较高的有效性和可靠性水平表明,使用设计的视标开发用于视力测试的有效且可靠的移动应用程序是可行的。该应用程序可由非医务人员随时随地用于视力筛查,以帮助提高公众对正确确定其视觉状态的意识和能力。