Repeatability of Orbscan III for anterior segment parameters in normal eyes.

作者信息

Sharma Mohita, Jain Neha, Koshy Ann S, Arora Vishal, Lalgudi Vaitheeswaran G

机构信息

Department of Cornea and Refractive Surgery, Tirupati Eye Centre, Noida, Uttar Pradesh, India.

Department of Cornea, Cataract and Refractive Surgery, Narayana Nethralaya Eye Hospital, Bangalore, Karnataka, India.

出版信息

Indian J Ophthalmol. 2020 Dec;68(12):2903-2905. doi: 10.4103/ijo.IJO_2856_20.

Abstract

PURPOSE

Assessment of repeatability of Orbscan III in measuring steep keratometry (sKm), flat keratometry (fKm), thinnest corneal thickness (TCT), anterior chamber depth (ACD), white-to-white diameter (WTW) and 3 mm and 5 mm zonal irregularity in healthy eyes.

METHODS

In this prospective observatioanl study 100 eyes of 50 participants underwent three consecutive scans on Orbscan III by a single examiner with a gap of 5 minutes between scans. Statistical analysis was performed to assess the repeatability of sKm, fKm, TCT, WTW, ACD and parameters of 3- and 5 mm irregularity. Within subject standard deviation (Sw), coefficient of variation (CoV) and test retest variability (TRT) was calculated for the commonly used parameters.

WERE THE MAIN OUTCOME MEASURES.: The repeatability of sKm, fKm, TCT, WTW, ACD, 3 mm and 5 mm zonal irregularity on Orbscan III.

RESULTS

Total of 50 patients (25 male and 25 female) who fulfilled inclusion and exclusion criteria were assessed. sKm, fKm, TCT and WTW were found to have a TRT of 0.31, 0.29. 21.5 and 0.27, respectively, and the ICC values were found to be over 0.9, denoting high repeatability. Repeat measurements of parameters of irregularity and ACD were found to be statistically similar, again denoting similarity between repeat measurements.

CONCLUSION

Orbscan III provides repeatable measurements of Sim-k, anterior chamber depth, thinnest corneal thickness, 3 mm and 5 mm zonal irregularity and white-to-white diameter in healthy eyes.

摘要

目的

评估Orbscan III在测量健康眼的陡峭角膜曲率(sKm)、平坦角膜曲率(fKm)、最薄角膜厚度(TCT)、前房深度(ACD)、白对白直径(WTW)以及3mm和5mm区域不规则度方面的可重复性。

方法

在这项前瞻性观察研究中,50名参与者的100只眼睛由一名检查者在Orbscan III上连续进行三次扫描,每次扫描间隔5分钟。进行统计分析以评估sKm、fKm、TCT、WTW、ACD以及3mm和5mm不规则度参数的可重复性。计算常用参数的受试者内标准差(Sw)、变异系数(CoV)和重测变异性(TRT)。

主要观察指标

Orbscan III上sKm、fKm、TCT、WTW、ACD、3mm和5mm区域不规则度的可重复性。

结果

共评估了50名符合纳入和排除标准的患者(25名男性和25名女性)。发现sKm、fKm、TCT和WTW的TRT分别为0.31、0.29、21.5和0.27,且ICC值超过0.9,表示具有高可重复性。发现不规则度参数和ACD的重复测量在统计学上相似,再次表明重复测量之间具有相似性。

结论

Orbscan III可对健康眼的Sim-k、前房深度、最薄角膜厚度、3mm和5mm区域不规则度以及白对白直径进行可重复测量。