J Cataract Refract Surg. 2012 Sep;38(9):1626-32. doi: 10.1016/j.jcrs.2012.04.031. Epub 2012 Jul 3.
To assess the repeatability of common measurements with the Sirius Scheimpflug-Placido topographer and Lenstar LS900 optical low-coherence reflectometry (OLCR) biometer and the limits of agreement (LoA) between the devices.
Eye Hospital of Wenzhou Medical College, Wenzhou, China.
Comparative evaluation of a diagnostic test or technology.
One randomly healthy eye of subjects was scanned 3 times with both devices. The parameters assessed were central corneal thickness (CCT), anterior chamber depth (ACD) from the corneal epithelium and from the endothelium, mean keratometry (K), and white-to-white (WTW) corneal diameter. The repeatability of scans was calculated using the within-subject standard deviation after 1-way analysis of variance was performed. The agreement between devices was assessed using the Bland-Altman LoA method, which equals the mean difference between devices ± 1.96 × standard deviation of the differences. The mean of 3 scans of each device was used to assess the LoA.
Forty subjects were evaluated. The repeatability of the Scheimpflug-Placido topographer and OLCR biometer was 3.10 μm and 3.32 μm for CCT, 0.04 mm and 0.05 mm for WTW corneal diameter, and 0.17 D and 0.10 D for mean K, respectively. The repeatability for both devices was 0.02 mm for the ACD from the corneal epithelium and the ACD from the corneal endothelium. On Bland-Altman LoA analysis, all parameters were within clinically acceptable limits.
Both devices had excellent repeatability for all parameters assessed. Good LoAs were found between the 2 devices, indicating they can be used interchangeably for the parameters assessed.
评估 Sirius Scheimpflug-Placido topography 与 Lenstar LS900 光学低相干反射计(OLCR)生物测量仪进行常见测量的重复性以及两种设备之间的一致性界限(LoA)。
温州医科大学附属眼视光医院,中国温州。
诊断测试或技术的比较评估。
使用两种设备对受试者的一只随机健康眼进行 3 次扫描。评估的参数包括中央角膜厚度(CCT)、角膜上皮和内皮的前房深度(ACD)、平均角膜曲率(K)和角膜直径(WTW)。在进行单因素方差分析后,使用个体内标准差计算扫描的重复性。使用 Bland-Altman LoA 方法评估设备之间的一致性,该方法等于设备之间的平均差异±1.96×差异标准差。使用每种设备的 3 次扫描的平均值评估 LoA。
评估了 40 名受试者。Scheimpflug-Placido topography 和 OLCR 生物测量仪的重复性分别为 CCT 3.10μm 和 3.32μm、WTW 角膜直径 0.04mm 和 0.05mm 以及平均 K 值 0.17D 和 0.10D。两种设备的 ACD 从角膜上皮和 ACD 从角膜内皮的重复性均为 0.02mm。在 Bland-Altman LoA 分析中,所有参数均在临床可接受范围内。
两种设备对评估的所有参数均具有出色的重复性。两种设备之间的 LoA 良好,表明可以互换使用评估的参数。