Suppr超能文献

三种锥形束 CT 设备和两种软件系统检测垂直根折的准确性。

Accuracy of three cone-beam CT devices and two software systems in the detection of vertical root fractures.

机构信息

Faculty of Dentistry, Federal University of Goiás, Goiânia, Goiás, Brazil.

Private Practioner, Goiânia, Goiás, Brazil.

出版信息

Dentomaxillofac Radiol. 2021 Mar 1;50(3):20200334. doi: 10.1259/dmfr.20200334. Epub 2020 Dec 1.

Abstract

OBJECTIVES

The aim of this study was to compare the accuracy of vertical root fracture (VRF) detection using three tomography devices and two software systems in teeth with different endodontic fillings.

METHODS

The sample consisted of 45 premolars divided into 3 groups: No filling (NF, n=15); Gutta percha (GP, n=15) and Metallic Post (MP, n=15). Cone-beam computed tomography (CBCT) images were acquired in Kodak 9000 3D, Orthopantomography 300 (OP300) and PreXion 3D devices, before and after induced root fractures. Two oral radiologists analyzed all images using InVivoDental and e-Vol DX software systems. The analysis was repeated after 15 days in 30% of the sample. Data analysis compared receiver operating characteristic (ROC) curves, as well the areas under the ROC curves. Accuracy, sensitivity, specificity, positive and negative predictive value were calculated according to each tomographic device and software. Intra- and interexaminer reliability were tested using the Kappa coefficient.

RESULTS

The highest accuracy was seen in the image set from the PreXion 3D, using InVivo (0.96) or e-Vol DX (0.92) in image analysis. The OP300 device presented a similar performance of the PreXion 3D in teeth with different endodontic fillings. When using e-Vol DX, the accuracy of Kodak 9000 3D improved from 0.62 to 0.74.

CONCLUSIONS

The PreXion 3D device is the most accurate when detecting VRF, with a performance similar to the OP300 in endodontic filled teeth. Kodak 9000 3D is indicated for teeth without fillings, with better accuracy using e-Vol DX software.

摘要

目的

本研究旨在比较三种体层摄影设备和两种软件系统在不同根管充填的牙齿中检测垂直根折(VRF)的准确性。

方法

样本由 45 颗前磨牙组成,分为三组:无充填(NF,n=15);牙胶(GP,n=15)和金属桩(MP,n=15)。在诱导根折前后,使用柯达 9000 3D、全景 300(OP300)和 PreXion 3D 设备获取锥形束 CT(CBCT)图像。两名口腔放射科医生使用 InVivoDental 和 e-Vol DX 软件系统分析所有图像。在 30%的样本中,分析在 15 天后重复进行。根据每个断层摄影设备和软件比较了接收者操作特征(ROC)曲线和 ROC 曲线下的面积。根据每个断层摄影设备和软件计算了准确性、敏感度、特异性、阳性和阴性预测值。使用 Kappa 系数测试了内部和内部检查者的可靠性。

结果

使用 InVivo(0.96)或 e-Vol DX(0.92)在图像分析中,PreXion 3D 图像组的准确性最高。在不同根管充填的牙齿中,OP300 设备的性能与 PreXion 3D 相似。当使用 e-Vol DX 时,Kodak 9000 3D 的准确性从 0.62 提高到 0.74。

结论

PreXion 3D 设备在检测 VRF 时最准确,在根管充填的牙齿中与 OP300 的性能相似。Kodak 9000 3D 适用于无充填的牙齿,使用 e-Vol DX 软件可提高准确性。

相似文献

引用本文的文献

本文引用的文献

7
Optimization of exposure parameters in dental cone beam computed tomography using a 3-step approach.采用三步法优化牙科锥形束计算机断层扫描中的曝光参数。
Oral Surg Oral Med Oral Pathol Oral Radiol. 2018 Dec;126(6):545-552. doi: 10.1016/j.oooo.2018.08.004. Epub 2018 Aug 20.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验