Urine Specific Gravity Effect on Total and Segmental Body Composition Validity of Multifrequency Bioelectrical Impedance Analysis Compared With Dual Energy X-Ray Absorptiometry.

机构信息

Flecks Rx Inc., Ridgway, Colorado; and.

Department of Kinesiology University of Wisconsin-Eau Claire, Eau Claire, Wisconsin.

出版信息

J Strength Cond Res. 2021 Feb 1;35(2):373-384. doi: 10.1519/JSC.0000000000003886.

Abstract

Fleck, SJ, Hayes, A, Stadler, G, Goesch, T, Goldammer, M, and Braun, S. Urine specific gravity effect on total and segmental body composition validity of multifrequency bioelectrical impedance analysis compared with dual energy x-ray absorptiometry. J Strength Cond Res 35(2): 373-384, 2021-The purposes were to compare body composition measures between a specific multifrequency bioelectrical impedance analyzer (InBody770) and dual-energy x-ray absorptiometry (DXA) and determine if hydration status within a specified range affected these measures. Methods included determining urine specific gravity before testing. Urine specific gravity needed to be within typical well-hydrated (n = 37), euhydrated (n = 45), or slightly dehydrated (n = 20) ranges. Segmental and total body composition measures were determined with the InBody770 and by DXA within the same testing session. Paired sample t-tests revealed significant differences (p < 0.005) between InBody770 and DXA for all body composition variables for all 3 hydration statuses, except for trunk fat-free mass (FFM) and trunk fat mass (FM) of the well-hydrated and euhydrated groups and right leg FM and trunk FFM of the slightly dehydrated group. For the total sample (n = 102), InBody770 significantly underestimated total body FM, right arm FFM, left arm FFM, right leg FFM, and left leg FFM with the range of underestimation being between 0.16 and 2.87 kg. The total body FFM by InBody770 was overestimated by 2.33 ± 2.80 kg or 3.6%. Bland-Altman plots supported these results. The major conclusions are that differences between the InBody770 and DXA segmental and total body FFM and FM are not significantly affected by hydration status in the range investigated, and the FM and FFM determined by the 2 devices are generally significantly different.

摘要

弗莱克,SJ,海耶斯,A,斯塔德勒,G,戈施,T,戈尔德默,M 和布劳恩,S. 尿比重对多频生物电阻抗分析与双能 X 射线吸收法相比的总身体成分和节段身体成分的影响。J 力量与体能研究 35(2):373-384,2021-目的是比较特定多频生物电阻抗分析仪(InBody770)和双能 X 射线吸收法(DXA)之间的身体成分测量值,并确定在特定范围内的水合状态是否会影响这些测量值。方法包括在测试前确定尿比重。尿比重需要在典型的良好水合(n = 37)、水合正常(n = 45)或轻度脱水(n = 20)范围内。在同一测试过程中,使用 InBody770 和 DXA 确定节段和总体身体成分测量值。配对样本 t 检验显示,对于所有 3 种水合状态的所有身体成分变量,InBody770 和 DXA 之间存在显著差异(p < 0.005),除了良好水合和水合正常组的躯干去脂体重(FFM)和躯干脂肪量(FM)以及轻度脱水组的右腿 FM 和躯干 FFM。对于总样本(n = 102),InBody770 显著低估了总身体 FM、右臂 FFM、左臂 FFM、右腿 FFM 和左腿 FFM,低估范围在 0.16 至 2.87 公斤之间。InBody770 测定的总身体 FFM 被高估了 2.33 ± 2.80 公斤或 3.6%。Bland-Altman 图支持这些结果。主要结论是,在研究范围内,InBody770 和 DXA 节段和总身体 FFM 和 FM 之间的差异不受水合状态的显著影响,并且这两种设备确定的 FM 和 FFM 通常存在显著差异。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验