三种不同设计的三维打印犬胸腰椎脊柱器械用动物专用钻头导向器在狗中的准确性和安全性:双侧和单侧设计。

Accuracy and safety of three-dimensionally printed animal-specific drill guides for thoracolumbar vertebral column instrumentation in dogs: Bilateral and unilateral designs.

机构信息

Division of Clinical Neurology, Vetsuisse Faculty, University of Bern, Bern, Switzerland.

Department of Surgical Sciences, School of Veterinary Medicine, University of Wisconsin- Madison, Madison, Wisconsin.

出版信息

Vet Surg. 2021 Feb;50(2):336-344. doi: 10.1111/vsu.13558. Epub 2020 Dec 19.

Abstract

OBJECTIVE

To evaluate the safety and accuracy of a unilateral three-dimensionally printed animal-specific drill guide (3DASDG) design for unilateral stabilization in the thoracolumbar vertebral column of dogs compared to a bilateral design.

STUDY DESIGN

Cadaveric study.

SAMPLE POPULATION

Fifty-two corridors in one canine cadaver.

METHODS

Two 3DASDG designs with 2 drilling tubes each were created from T8 to L7 vertebrae. Fifty-two corridors were drilled on the right and the left sides by using unilateral and bilateral designs, respectively. Planned and postoperative trajectories (entry point, exit point, angle) were compared to establish the accuracy. Statistical analysis was used for accuracy comparison between designs. Safety was evaluated by using Zdichavsky classification.

RESULTS

Unilateral and bilateral drill guide designs were not different for entry point and angle deviations; however, they were different for the exit point deviations. Two corridors breached outside the vertebra. For all guides, mean entry and exit point deviations were less than 1 and 2 mm, respectively. The maximum entry or exit point deviation in both groups was 4.7 mm. The mean angle deviation was <3.5°, and the maximum angle deviation was 9.3°.

CONCLUSION

No difference was detected in accuracy of entry points and angle deviations between drill guide designs tested in normal vertebrae. The technique was classified as highly safe.

CLINICAL SIGNIFICANCE

A unilateral drill guide design may be a safe alternative to bilateral guides for unilateral stabilization of the thoracolumbar vertebral column in dogs.

摘要

目的

评估单侧三维打印动物专用钻导板(3DASDG)设计用于犬胸腰椎单侧稳定的安全性和准确性,与双侧设计相比。

研究设计

尸体研究。

样本人群

一只犬的 52 个椎管。

方法

从 T8 到 L7 椎骨分别为双侧设计创建了两个带有 2 个钻管的 3DASDG 设计。分别使用单侧和双侧设计在右侧和左侧钻 52 个椎管。比较计划和术后轨迹(进入点、出口点、角度)以确定准确性。使用 Zdichavsky 分类法评估安全性。

结果

单侧和双侧钻导板设计在进入点和角度偏差方面没有差异,但在出口点偏差方面有差异。有两个椎管穿出椎体外。对于所有引导器,平均进入和退出点偏差均小于 1 和 2mm,分别。两组的最大进入或退出点偏差均为 4.7mm。平均角度偏差<3.5°,最大角度偏差为 9.3°。

结论

在正常椎骨中,测试的钻导板设计在进入点和角度偏差的准确性方面没有差异。该技术被归类为高度安全。

临床意义

单侧钻导板设计可能是犬胸腰椎单侧稳定的双侧导板的安全替代方案。