两点 Dixon T2 加权成像在甲状腺相关眼病中的应用:与常规脂肪抑制成像在脂肪抑制质量和分期性能方面的比较。

Usefulness of two-point Dixon T2-weighted imaging in thyroid-associated ophthalmopathy: comparison with conventional fat saturation imaging in fat suppression quality and staging performance.

机构信息

Department of Radiology, The First Affiliated Hospital of Nanjing Medical University, Nanjing, China.

Department of Endocrinology, The First Affiliated Hospital of Nanjing Medical University, Nanjing, China.

出版信息

Br J Radiol. 2021 Feb 1;94(1118):20200884. doi: 10.1259/bjr.20200884. Epub 2020 Dec 22.

Abstract

OBJECTIVE

To compare the two-point Dixon weighted imaging (WI) with conventional fat-sat WI in fat suppression (FS) quality and staging performance for patients with TAO.

METHODS

We enrolled 37 thyroid-associated ophthalmopathy (TAO) patients and 15 healthy controls who underwent both coronal two-point Dixon and fat-sat WI. Qualitative (overall imaging quality, FS uniformity) and quantitative [signal intensity ratio of extraocular muscle (EOM-SIR)] parameters were assessed between the two-point Dixon WI and fat-sat WI. Additionally, water fraction of intraorbital fat (IF-WF) was measured on Dixon image. Dixon-EOM-SIR, Fat-sat-EOM-SIR and Dixon-IF-WF values were compared between active and inactive TAO groups, and the diagnostic efficiency for the active phase were evaluated.

RESULTS

Two-point Dixon WI showed significantly higher overall image quality score, FS uniformity score as well as EOM-SIR value than fat-sat WI in both TAO and control groups (all < 0.05). Active TAOs had significantly higher Dixon-EOM-SIR ( < 0.001), Fat-sat-EOM-SIR ( < 0.001) and Dixon-IF-WF ( = 0.001) than inactive TAOs. ROC curves analyses indicated that Dixon-EOM-SIR ≥3.32 alone demonstrated the highest staging sensitivity (75.0%). When integrating Dixon-EOM-SIR ≥3.32 and Dixon-IF-WF ≥0.09, improved staging efficiency and specificity could be achieved (area under the curve, 0.872; specificity, 97.1%).

CONCLUSION

Compared with conventional fat-sat technique, two-point Dixon WI offers better image quality, as well as improved staging sensitivity and specificity for TAO. Dixon WI is suggested to be used to evaluate the patients with TAO in clinical practice.

ADVANCES IN KNOWLEDGE

Two-point Dixon WI offers better image quality than fat-sat WI. Dixon-EOM-SIR alone demonstrated the highest staging sensitivity. Combining with Dixon-IF-WF showed improved staging efficiency and specificity. Dixon WI is suggested to be used to evaluate TAO patients in clinical practice.

摘要

目的

比较两点 Dixon 加权成像(WI)与常规脂肪饱和 WI 在甲状腺相关眼病(TAO)患者脂肪抑制(FS)质量和分期表现方面的差异。

方法

我们纳入了 37 例 TAO 患者和 15 例健康对照者,他们均接受了冠状位两点 Dixon 和脂肪饱和 WI 检查。评估了两点 Dixon WI 和脂肪饱和 WI 的定性(整体成像质量、FS 均匀性)和定量[眼外肌(EOM)信号强度比(EOM-SIR)]参数。此外,还在 Dixon 图像上测量了眼眶内脂肪的水分数(IF-WF)。比较了活动期和非活动期 TAO 组之间 Dixon-EOM-SIR、Fat-sat-EOM-SIR 和 Dixon-IF-WF 值,并评估了其对活动期的诊断效能。

结果

两点 Dixon WI 在 TAO 患者和对照组中均显示出明显更高的整体图像质量评分、FS 均匀性评分和 EOM-SIR 值(均<0.05)。活动期 TAO 的 Dixon-EOM-SIR(<0.001)、Fat-sat-EOM-SIR(<0.001)和 Dixon-IF-WF(=0.001)均显著高于非活动期 TAO。ROC 曲线分析表明,Dixon-EOM-SIR≥3.32 单独具有最高的分期敏感性(75.0%)。当整合 Dixon-EOM-SIR≥3.32 和 Dixon-IF-WF≥0.09 时,可以提高分期的敏感性和特异性(曲线下面积,0.872;特异性,97.1%)。

结论

与常规脂肪饱和技术相比,两点 Dixon WI 为 TAO 提供了更好的图像质量,以及更高的分期敏感性和特异性。建议在临床实践中使用 Dixon WI 评估 TAO 患者。

知识进展

两点 Dixon WI 提供了比脂肪饱和 WI 更好的图像质量。Dixon-EOM-SIR 单独具有最高的分期敏感性。结合 Dixon-IF-WF 可提高分期的敏感性和特异性。建议在临床实践中使用 Dixon WI 评估 TAO 患者。

相似文献

引用本文的文献

3
Application progress of magnetic resonance imaging in thyroid associated ophthalmopathy.磁共振成像在甲状腺相关性眼病中的应用进展
Front Endocrinol (Lausanne). 2025 Jun 26;16:1537957. doi: 10.3389/fendo.2025.1537957. eCollection 2025.

本文引用的文献

2
Imaging Investigation of Thyroid Eye Disease.甲状腺眼病的影像学检查
Ophthalmic Plast Reconstr Surg. 2018 Jul/Aug;34(4S Suppl 1):S41-S51. doi: 10.1097/IOP.0000000000001139.
6
Thyroid-associated Ophthalmopathy.甲状腺相关性眼病
Turk J Ophthalmol. 2017 Apr;47(2):94-105. doi: 10.4274/tjo.80688. Epub 2017 Apr 1.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验