Direct monitoring of microvascular anastomoses with the 20-MHz ultrasonic Doppler probe: an experimental and clinical study.

作者信息

Swartz W M, Jones N F, Cherup L, Klein A

机构信息

Division of Plastic and Reconstructive Surgery, University of Pittsburgh, Pa.

出版信息

Plast Reconstr Surg. 1988 Feb;81(2):149-61. doi: 10.1097/00006534-198802000-00001.

Abstract

The 20-MHz ultrasonic Doppler probe was used to determine its efficacy as a continuous monitoring technique for microvascular anastomoses. A 1-mm2 piezoelectric crystal embedded in a soft silicone sleeve was sutured directly to the blood vessel distal to the anastomosis. Using the dog femoral artery, simultaneous measurement of velocity and blood flow with an electromagnetic flowmeter established a direct correlation between flow and velocity with a 14 percent error at maximum flow and an 18 percent error at minimum flow conditions. The probe was then implanted in the rabbit femoral artery for 1 week (n = 3) and 4 weeks (N = 6), demonstrating that a continuous tracing could be obtained without injury to the vessel. Our clinical study included 63 patients undergoing free-tissue transfers monitored with the implantable probe for 7 to 29 days (average 10.5 days). Twenty-three flaps were buried. Two patients experienced loss of arterial tracing due to malfunction of the probe (3 percent). Three patients had a venous thrombosis with a present arterial tracing. There were no flap failures per se. All probes were removed without mishap, and there were no complications related to the probe. We conclude that the 20-MHz ultrasonic Doppler probe holds promise as a useful monitoring method.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验