Angle Orthod. 2021 May 1;91(3):336-342. doi: 10.2319/050120-375.1.
To compare the primary stability of miniscrews after repeated cycles of insertion through insertion torque (IT) measurements and resonance frequency analysis (RFA).
Sixty titanium miniscrews were divided into two groups according to the insertion protocol: one with predrilled sites and the other self-drilled into porcine iliac crest bone specimens. Each group had three cycles of reinsertion. After each insertion, the IT and RFA were measured. The IT was measured by using a torque meter, and the RFA was measured using the Osstell ISQ device. A total of five miniscrews of each group were selected for sequential assessment of the morphology of their tip and threads using scanning electron microscopy after each insertion cycle.
No statistically significant differences were found in the IT values of miniscrews reinserted up to three times in the group with predrilled surgical sites. The IT value increased significantly with the number of reinsertions in the self-drilled group. The RFA value decreased as the number of insertions increased in both groups.
Under the conditions of this in vitro study, reinserting miniscrews deteriorates the integrity of their tip and thread. Reinsertion should be discouraged particularly when insertion sites are not predrilled.
通过测量插入扭矩 (IT) 和共振频率分析 (RFA) 比较微型种植体在重复插入循环后的初始稳定性。
将 60 颗钛制微型种植体根据插入方案分为两组:一组为预钻部位,另一组为自行钻入猪髂嵴骨标本。每组进行三次重复插入。每次插入后测量 IT 和 RFA。使用扭矩计测量 IT,使用 Osstell ISQ 设备测量 RFA。每组选择五个微型种植体,在每次插入循环后使用扫描电子显微镜对其尖端和螺纹的形态进行连续评估。
在预钻手术部位的组中,微型种植体重复插入三次的 IT 值没有统计学差异。在自行钻入组中,IT 值随插入次数的增加而显著增加。在两组中,RFA 值随着插入次数的增加而降低。
在这项体外研究的条件下,重复插入微型种植体破坏了其尖端和螺纹的完整性。特别是在未预钻插入部位时,应避免重复插入。