一种使用 Osstell ISQ 设备测量正畸微种植体稳定性的新方法。

A novel technique for measurement of orthodontic mini-implant stability using the Osstell ISQ device.

出版信息

Angle Orthod. 2019 Mar;89(2):284-291. doi: 10.2319/011518-46.1. Epub 2018 Nov 1.

Abstract

OBJECTIVES

To develop and validate a method for application of the Osstell ISQ device in the assessment of mini-implant stability.

MATERIALS AND METHODS

An adaptor was developed for attachment of Osstell's SmartPeg onto a variety of orthodontic mini-implants. For validation of the adaptor, Benefit mini-implants were inserted into bone blocks that mimicked different stability conditions. The Osstell device was used to assess mini-implant stability with the adaptor (test measurement) and conventional SmartPeg attachment (gold-standard measurement). Implant stability quotient (ISQ) values were assessed for agreement, repeatability, and reproducibility.

RESULTS

Strong positive correlations were found between ISQ values obtained using the novel adaptor and the conventional attachment. Repeatability and reproducibility of ISQ values with the adaptor were similar to those obtained with the conventional attachment.

CONCLUSIONS

A method was developed and validated to assess the stability of orthodontic mini-implants using the Osstell system. The novel mini-implant adaptor provided repeatable and reproducible measurements of mini-implant stability, which agreed with those obtained using a conventional SmartPeg attachment. This adaptor permits noninvasive stability assessment of various designs of mini-implants, most of which are incompatible with the conventional SmartPeg attachment.

摘要

目的

开发并验证一种应用 Osstell ISQ 设备评估微型种植体稳定性的方法。

材料和方法

为将 Osstell 的 SmartPeg 连接到各种正畸微型种植体上,开发了一种适配器。为了验证适配器,将 Benefit 微型种植体插入模仿不同稳定性条件的骨块中。使用 Osstell 设备通过适配器(测试测量)和传统的 SmartPeg 连接(金标准测量)评估微型种植体稳定性。评估了植入物稳定性指数(ISQ)值的一致性、重复性和再现性。

结果

使用新型适配器获得的 ISQ 值与传统附件获得的 ISQ 值之间存在强正相关。使用适配器获得的 ISQ 值的重复性和再现性与使用传统附件获得的 ISQ 值相似。

结论

开发并验证了一种使用 Osstell 系统评估正畸微型种植体稳定性的方法。新型微型种植体适配器可提供微型种植体稳定性的可重复和可再现的测量值,与使用传统 SmartPeg 附件获得的测量值一致。该适配器允许对各种设计的微型种植体进行非侵入性稳定性评估,其中大多数与传统的 SmartPeg 附件不兼容。

相似文献

1
A novel technique for measurement of orthodontic mini-implant stability using the Osstell ISQ device.
Angle Orthod. 2019 Mar;89(2):284-291. doi: 10.2319/011518-46.1. Epub 2018 Nov 1.
3
Comparative analysis of 2 resonance frequency measurement devices: Osstell Mentor and Osstell ISQ.
Implant Dent. 2014 Jun;23(3):351-6. doi: 10.1097/ID.0000000000000072.
4
Customized SmartPeg for measurement of resonance frequency of mini dental implants.
Int J Implant Dent. 2017 Dec;3(1):4. doi: 10.1186/s40729-017-0066-6. Epub 2017 Feb 1.
5
A novel device for resonance frequency assessment of one-piece implants.
Int J Oral Maxillofac Implants. 2012 May-Jun;27(3):523-7.
6
Intra- and inter-operator concordance of the resonance frequency analysis. A cross-sectional and prospective clinical study.
Clin Oral Investig. 2022 Nov;26(11):6521-6530. doi: 10.1007/s00784-022-04601-y. Epub 2022 Jul 8.
7
Comparison of Reliability of 3 Resonance Frequency Analysis Devices: An In Vitro Study.
J Oral Implantol. 2022 Feb 1;48(1):9-14. doi: 10.1563/aaid-joi-D-20-00139.
10
Measurement of mini-implant stability using resonance frequency analysis.
Angle Orthod. 2013 Mar;83(2):230-8. doi: 10.2319/043012-354.1. Epub 2012 Aug 29.

引用本文的文献

1
Optimization Analysis of Two-Factor Continuous Variable between Thread Depth and Pitch of Microimplant under Toque Force.
Comput Math Methods Med. 2022 Jun 20;2022:2119534. doi: 10.1155/2022/2119534. eCollection 2022.
2
Comparison of the primary stability of orthodontic miniscrews after repeated insertion cycles.
Angle Orthod. 2021 May 1;91(3):336-342. doi: 10.2319/050120-375.1.

本文引用的文献

1
A Comparison of the Mechanical Measures Used for Assessing Orthodontic Mini-Implant Stability.
Implant Dent. 2017 Apr;26(2):225-231. doi: 10.1097/ID.0000000000000514.
2
Assessment of Osstell™ and Periotest® systems in measuring dental implant stability (in vitro study).
Saudi Dent J. 2011 Jan;23(1):17-21. doi: 10.1016/j.sdentj.2010.09.003. Epub 2010 Nov 13.
3
Measurement of mini-implant stability using resonance frequency analysis.
Angle Orthod. 2013 Mar;83(2):230-8. doi: 10.2319/043012-354.1. Epub 2012 Aug 29.
4
Assessment of miniscrew implant stability by resonance frequency analysis: a study in human cadavers.
J Oral Maxillofac Surg. 2010 Nov;68(11):2682-9. doi: 10.1016/j.joms.2010.05.083.
5
Miniscrews in orthodontic treatment: review and analysis of published clinical trials.
Am J Orthod Dentofacial Orthop. 2010 Jan;137(1):108-13. doi: 10.1016/j.ajodo.2008.01.027.
7
Soft bone primary stability of 3 different miniscrews for orthodontic anchorage: a resonance frequency investigation.
Am J Orthod Dentofacial Orthop. 2009 May;135(5):642-8. doi: 10.1016/j.ajodo.2007.05.025.
8
Mini-implants in orthodontics: a systematic review of the literature.
Am J Orthod Dentofacial Orthop. 2009 May;135(5):564.e1-19; discussion 564-5. doi: 10.1016/j.ajodo.2008.09.026.
10
Influence of defect depth on resonance frequency analysis and insertion torque values for implants placed in fresh extraction sockets: a human cadaver study.
Clin Implant Dent Relat Res. 2009 Mar;11(1):52-8. doi: 10.1111/j.1708-8208.2008.00095.x. Epub 2008 Apr 1.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验