CT dose management of adult patients with unknown body weight using an effective diameter.

机构信息

Department of Radiological Technology, Kurashiki Central Hospital, 1-1-1 Miwa, Kurashiki, Okayama, 710-8602, Japan; Division of Health Sciences, Graduate School of Medical Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, 920-0942, Japan.

Division of Health Sciences, Graduate School of Medical Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, 920-0942, Japan.

出版信息

Eur J Radiol. 2021 Feb;135:109483. doi: 10.1016/j.ejrad.2020.109483. Epub 2020 Dec 18.

Abstract

PURPOSE

To assess the usefulness of effective diameter (D) for CT dose management of adult patients with unknown body weight.

METHODS

A total of 642 adult patients whose height and weight had been measured before CT examination (chest CT using Aquilion Prime SP, 428 patients; chest CT using Biograph mCT, 100 patients; and abdominal CT using Aquilion Prime SP, 114 patients) were retrospectively examined between April 2018 and September 2019. The D was automatically calculated from the lateral diameter on a CT localizer radiograph by a dose management software (Radimetrics). In order to determine the correlation between body weight and D, we compared volume CT dose index and dose length product between patients with body weight between 50 and 70 kg and those with D equivalent to body weight between 50 and 70 kg. Correlation analysis was performed by Pearson's product-moment correlation, and statistical analyses were performed by using t-test.

RESULTS

The correlation coefficient values between body weight and D were 0.920 for chest CT using Aquilion Prime SP, 0.929 for chest CT using Biograph mCT, and 0.805 for abdominal CT using Aquilion Prime SP. In both chest and abdominal CT scans, there were no significant differences in volume CT dose index and dose length product between patients with body weight between 50 and 70 kg and those with D equivalent to body weight between 50 and 70 kg.

CONCLUSIONS

The D may be useful as a somatometric parameter for CT dose management of adult patients with unknown body weight.

摘要

目的

评估有效直径(D)在管理未知体重成人患者 CT 剂量中的作用。

方法

回顾性分析 2018 年 4 月至 2019 年 9 月期间行 CT 检查(胸部 CT:Aquilion Prime SP 428 例,Biograph mCT100 例;腹部 CT:Aquilion Prime SP114 例)且身高和体重可测量的 642 例成年患者资料。D 由剂量管理软件(Radimetrics)自动从 CT 定位片的侧径计算得出。为确定体重与 D 之间的相关性,我们比较了体重 50-70kg 患者与 D 相当于体重 50-70kg 患者的容积 CT 剂量指数和剂量长度乘积。采用 Pearson 积差相关分析进行相关性分析,采用 t 检验进行统计学分析。

结果

胸部 CT(Aquilion Prime SP)、胸部 CT(Biograph mCT)和腹部 CT(Aquilion Prime SP)中体重与 D 的相关系数值分别为 0.920、0.929 和 0.805。在胸部和腹部 CT 扫描中,体重 50-70kg 患者与 D 相当于体重 50-70kg 患者的容积 CT 剂量指数和剂量长度乘积无显著差异。

结论

D 可作为管理未知体重成年患者 CT 剂量的体尺参数。