脉络膜毛细血管的扫频源光相干断层血管造影成像。

SWEPT-SOURCE OPTICAL COHERENCE TOMOGRAPHY ANGIOGRAPHY IMAGING OF THE CHORIOCAPILLARIS.

机构信息

Vitreous Retina Macula Consultants of New York, New York.

出版信息

Retina. 2021 Jul 1;41(7):1373-1378. doi: 10.1097/IAE.0000000000003109.

Abstract

PURPOSE

To analyze swept-source optical coherence angiography images acquired at different depths above and below the default location of the Zeiss PLEX Elite 9000.

METHODS

Normal eyes of subjects in their 20s and 30s were evaluated. Angiographic slab images were taken at the default location of 29 µm to 49 µm below the retinal pigment epithelium and 21 µm to 41 µm through 52 µm to 72 µm below in steps. The images were processed using the projection removal function from the device's software. Raw images were evaluated, as were images that underwent a published compensation technique that adjusts for light penetration to the sampled layer.

RESULTS

Eleven eyes of 11 subjects were evaluated for the uncompensated and the compensated sets with the projection removal function turned off and on. The default location, 29 µm to 49 µm below the retinal pigment epithelium, showed a granular choriocapillaris appearance. This appearance remained in all slabs from each group, differing slightly throughout depth. The projection removal function modified the grayscale values and diminished projection from overlaying retinal vessels. The compensation technique altered the appearance of flow deficits, and the changes induced by it were more evident on the images were the projection removal function was turned on.

CONCLUSION

Flow images in swept-source optical coherence angiography of the choriocapillaris from varying levels are similar in appearance, suggesting projection from the choriocapillaris is important in image formation, although layers of vessels in the inner choroid may contribute by various amounts. A model explaining the prominent projection artifacts observed in the choroid with swept-source optical coherence angiography imaging is presented.

摘要

目的

分析在蔡司 PLEX Elite 9000 的默认位置上下不同深度处采集的扫频源光相干血管造影图像。

方法

评估了 20 多岁和 30 多岁的受试者的正常眼睛。在视网膜色素上皮下方 29μm 至 49μm 和下方 21μm 至 41μm 的位置以步长穿过 52μm 至 72μm 采集血管造影板图像。使用设备软件中的投影去除功能处理图像。评估了原始图像和经过光穿透到采样层的已发布补偿技术调整的图像。

结果

使用关闭和打开投影去除功能的未补偿和补偿集评估了 11 只眼睛的 11 名受试者。在视网膜色素上皮下方 29μm 至 49μm 的默认位置显示出颗粒状脉络膜毛细血管外观。这种外观在每组的所有板中均保持不变,在深度上略有不同。投影去除功能改变了灰度值,并减少了来自重叠视网膜血管的投影。补偿技术改变了血流不足的外观,并且当投影去除功能打开时,它引起的变化更为明显。

结论

从不同水平获取的脉络膜毛细血管的扫频源光相干血管造影图像中的血流图像在外观上相似,表明来自脉络膜毛细血管的投影对图像形成很重要,尽管脉络膜内层的血管层可能以不同的量贡献。提出了一个解释扫频源光相干血管造影成像中脉络膜中观察到的突出投影伪影的模型。