通过连续监测植入物负载评估骨折愈合的临床可行性。

Clinical feasibility of fracture healing assessment through continuous monitoring of implant load.

机构信息

AO Research Institute Davos, Davos, Switzerland.

BG Trauma Center Tübingen, Eberhard Karls University Tübingen, Tübingen, Germany.

出版信息

J Biomech. 2021 Feb 12;116:110188. doi: 10.1016/j.jbiomech.2020.110188. Epub 2020 Dec 24.

Abstract

Current fracture fixation follow-up is based on subjective radiological and clinical examination. Efforts to objectify the procedure have been undertaken since decades. Assessment of implant load as an indirect predictor of callus maturity has so far failed to enter clinical routine due to limited practicability, technical obstacles and its snap-shot nature. We recently introduced the concept of continuous implant load monitoring to aid in diagnosing fracture healing progression. This study aimed at investigating the feasibility of the system in a clinical context. Ten patients treated with Taylor-Spatial-Frame external fixators following pathological tibia fractures were equipped with a Fracture Monitor device attached to a fixator-strut and were monitored until hardware removal. Two patients were excluded due to technical issues. Implant load and fracture activity was continuously and autonomously measured for 139 ± 89 days (mean ± SD). Data was wirelessly collected with consumer smartphones. Relative implant load initially rose for 34.1 ± 22.2 days and finally declined to a level of 45.0 ± 33.8% of the maximum implant load. In five patients the load dropped below 50% of the maximum load. These patients underwent hardware removal according to the clinical assessment. In three patients, whose external fixators were exchanged to internal fixation at the end of the study, implant load did not drop below the 50% margin. The continuous measurement principle allows resolving implant load progression and appears indicative for the bone healing status. Data can be acquired in a homecare setting and is believed to provide valuable information to support timely healing assessment and enable patient specific after-care.

摘要

目前的骨折固定随访基于主观的影像学和临床检查。几十年来,人们一直致力于使该过程客观化。评估作为骨痂成熟度间接预测指标的植入物负载,由于实用性有限、技术障碍以及其快照性质,尚未进入临床常规。我们最近提出了连续植入物负载监测的概念,以帮助诊断骨折愈合进展。本研究旨在调查该系统在临床环境中的可行性。10 名接受病理性胫骨骨折 Taylor-Spatial-Frame 外固定架治疗的患者配备了一个固定器支柱上的 Fracture Monitor 设备,并进行监测直到去除内固定物。由于技术问题,有两名患者被排除在外。植入物负载和骨折活动被连续和自主地测量了 139±89 天(平均值±标准差)。数据通过消费类智能手机无线收集。相对植入物负载最初升高 34.1±22.2 天,最后降至最大植入物负载的 45.0±33.8%。在五名患者中,负载降至最大负载的 50%以下。这些患者根据临床评估去除了内固定物。在三名患者中,他们的外固定器在研究结束时被更换为内固定,植入物负载没有降至 50%的边界以下。连续测量原理允许解析植入物负载的进展,并且似乎对骨愈合状态具有指示性。数据可以在家庭护理环境中获得,据信可以提供有价值的信息,以支持及时的愈合评估并实现患者特定的康复护理。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验